Sensor data format, Calibration numbers, Global water instrumentation – Xylem SIT65 SATELLITE INTERNET TELEMETRY User Manual

Page 13

Advertising
background image

Global Water Instrumentation

13

Sensor Data Format

Analog sensor data is transmitted as a 10-bit number in the range of 0-1023. A

calibration equation is applied to the raw data number at the database to scale the

sensor reading in the appropriate range and engineering units. Digital input

accumulators count input events from things like tipping bucket rain gauges and flow

monitors. The accumulator is transmitted as a 11-bit number in the range of 0-2047

and calibration factors are also used to scale the data. Once reaching 2047, the

accumulator counter rolls back to zero with the next count. Digital status inputs

transmit all channel data in a single status register with 1-bit for each channel, one or

zero, true/false, on/off, open/closed, etc. Digital control outputs also have their

status transmitted in a single register with 1-bit for each output channel.

Because digital status inputs and control outputs only need one data bit to transmit

their state, these channels can be combined into one data value. The state of up to

4 status inputs are combined into one status register in binary format, channel one

shown by bit one, channel two with bit two, etc. In the same manner, the current

state of the control outputs is transmitted in a single status register. Register

contents returned as 0011 (binary) shows channels 1 and 2 High (ON), and channels

3 and 4 Low (OFF). Features in the databases allow the mapping of these bits to

text labels.

Calibration Numbers

Data values are transmitted to the hosting databases as raw data numbers. This

number is scaled to the desired engineering units (EU) by first multiplying it with a

scaling factor (Multiplier), then adding an offset (Adder). These numbers are the

coefficients in a straight line equation: EU = Adder + (Multiplier * RawData). These

calibration numbers are entered into the databases and the scaling takes place as

the data comes in. Thus, all the calibration is done at the database, independent of

the field station. There is an Excel spreadsheet calculator for finding these

calibration factors that can be downloaded from our website at globalw.com/support.

Some examples are also given in Appendix B.

Another important feature is available for calculating data. When the raw data has

been adjusted to the proper engineering units using the straight line equation

discussed above, the database calibrations allows for additional mathematical

operations. For example; these equations can be flume or weir equations for

Advertising