IMPORTANT:

BOTH THE WIND ALARM READOUT AND THE WIND

SPEED SENSOR (GENERATOR) MUST BE RETURNED TO
MAXIMUM, INC. ON AN ANNUAL BASIS FOR CALIBRATION
CHECK. FAILURE TO RETURN BOTH COMPONENTS WILL VOID
THE MANUFACTURERS LIMITED 5-YEAR WARRANTY, AND ALSO
SIGNIFY THAT THE INSTRUMENT OWNER/USER AGREES TO
HOLD HARMLESS MAXIMUM, INC. SHOULD A DEVICE TO WHICH
THE WIND ALARM IS CONNECTED, CAUSE PERSONAL HARM OR
PROPERTY DAMAGE.

Trouble Shooting

If the read out is, blank (no lights).

Check power source; make sure that there is 12VDC supplied.

Check Polarity, make sure positive and negative leads were connected according to
instructions.

Check sensor connections; Make sure the wiring sequence for wind speed sensor is
correct.

If there is no Wind Speed (reads 00).

Check continuity of the cable.

Check sensor connections; Make sure the wiring sequence for wind speed sensor is
correct.

Disconnect the wires marked #1, #2, and #3 from the terminal strip. With a short
jumper wire short (touch together) terminals #1 and #2 together several times. This
will make the read-out show a random number for wind speed. If it does instrument is
OK, and the problem is in the sensor. If it still shows “00” the problem is in the read-
out.

To Test Alarm

Set the set points to “00” MPH. This will activate both alarms.

All Maximum instruments are covered by a 5-year limited warranty. Please see
warranty card for further details. If you have any questions please feel free to contact