Installation instructions, Recommended installation operating instructions – Pump Solutions Group Neptune Calibration Columns User Manual

Page 2

Advertising
background image

Installation Instructions

The recommended installation is shown below:

1. Install the Calibration Column in the suction line

of the Chemical Feed Pump. The column should be
installed in a vertical position.

2. Two (2) Ball Valves are required and must be

installed as shown in the adjacent drawing. Valves
are not included.

3. The column should be installed so it will be filled

regardless of the level of the tank. It is recommended
that the top connection of the column be connected
to a vent tube that is as high as the liquid level in the
tank to prevent accidental spillage in the event of a
valve malfunction or if the isolation valve is acciden-
tally left open. Do not install any valve in this overflow
line, as the test tube must be vented to atmosphere
at all times.

CAUTION: Top cap is not glued to allow cleaning. Top
cap must be glued when vent is piped back to tank.

CAUTION: If the vent is not piped back to the tank
a “U” tube should be installed on the top of the col-
umn to prevent accidental splash-up for operator pro-
tection. Remember, the head in a tall tank can cause
the column to fill rapidly resulting in a squirt from the
small vent connection on the top of the column.

NOTE: When viewing the pump’s performance on a
calibration cylinder, each stroke of the pump should
steadily draw the liquid down to a dead stop and
then drawdown again. “Bouncing” in the column at
the end of the suction stroke is an indication of dirt
or clogging in the valves or possibly wear.

CAUTION —
IMPORTANT INSTALLATION NOTES:

a. This unit must be vented to atmosphere when in use.

b. Never use this device on discharge side of pump.

c. This device is intended for verification of metering

pump flow rate only.

d. Maximum Pressure = 25 psig

Maximum Temperature = 140°F

e. Not suitable for all chemicals. Consult Factory. Shielded

glass models available for acids and strong chemicals.

Recommended Installation

Operating Instructions

A stopwatch or wristwatch with sweep second hand
is required.

1. With the pump operating normally, Valve A is open

and Valve B is closed. Open Valve B slightly so the
column fills slowly with liquid.

2. When the liquid level reaches the zero division mark,

close Valve A (any division mark may be used as a
starting point, however, by filling to the zero mark,
you may measure over a longer period of time, ena-
bling you to get a better calibration). Do not fill above
the zero division mark unless top cap is glued and
vent is piped to tank.

3. Allow liquid level to drop for a period of at least

30 seconds and note the mark on the column to
which the level has dropped. Re-open Valve A.

4. Each column has two scales. The left hand scale

shows the volume of the column in ml. Dividing the
volume pumped by the time to pump gives you the
flow rate. The right hand scale is direct reading in
gallons per hour for a 1 minute test. If the test time is
exactly 1 minute, the drop in the column indicates the
flow rate. Double the number on the right hand scale
for a 30 second test.

5. When the gauge is not in use, Valve B must be closed.

6. Valve A must remain open at all times except when

testing the pump rate.

NEPTUNE

PUMP

PUMP

SUCTION

VALVE A

RESERVOIR

OVERFLOW PROTECTION LINE

CALIBRATION COLUMN

VALVE B

© 2008 Neptune Chemical Pump Company

All Rights Reserved

Bulletin CC-08

P/N ZL107028

Printed in U.S.A.

SOLD BY:

P.O. Box 247 • Lansdale, PA 19446-0247
Tel: 215-699-8700 • Fax: 215-699-0370
Toll-Free Tel: 1-888-3NEPTUNE (1-888-363-7886)
Toll-Free Fax: 1-800-255-4017
Web Site: http://www.neptune1.com
E-mail: [email protected]

Advertising