AAEON ACP-5212 User Manual

Page 6

Advertising
background image

M u l t i - T o u c h P a n e l P C

A C P - 5 2 1 2

v

14. If any of the following situations arises, get the equipment

checked by service personnel:

a. The power cord or plug is damaged.

b. Liquid has penetrated into the equipment.

c. The equipment has been exposed to moisture.

d. The equipment does not work well, or you cannot get it to

work according to the users manual.

e. The equipment has been dropped and damaged.

f. The equipment has obvious signs of breakage.

15.

DO NOT LEAVE THIS EQUIPMENT IN AN UNCONTROLLED
ENVIRONMENT WHERE THE STORAGE TEMPERATURE IS
BELOW -20° C (-4°F) OR ABOVE 60° C (140° F). IT MAY
DAMAGE THE EQUIPMENT

.


16. External equipment intended for connection to signal

input/output or other connectors, shall comply with relevant
UL / IEC standard (e.g. UL 1950 for IT equipment and
UL 60601-1 / IEC 60601 series for systems – shall comply with
the standard IEC 60601-1-1, Safety requirements for medical
electrical systems. Equipment not complying with
UL 60601-1 shall be kept outside the patient environment, as
defined in the standard.

Caution:

It may cause the danger of explosion if battery is incorrectly
replaced. Replace only with same or equivalent type
recommended by the manufacturer.

Advertising