Definitions

• All specifications and characteristics apply under the following conditions, unless otherwise stated:
• After 15 minutes of warm-up time, where the instrument is left in the ON state.
• Temperature range is 23 °C ± 5 °C.
• All specifications apply when using internal reference.
• All specifications subject to change without notice. Please visit www.anritsu.com for most current datasheet.
• Typical performance is the measured performance of an average unit.
• Recommended calibration cycle is 12 months.

Frequency

Frequency Range: 500 kHz to 1.6 GHz (500 kHz to 6.0 GHz with Option 16)
Frequency Accuracy: 2.5 ppm
Frequency Resolution: 1 Hz

Frequency Range

High Port Power

Low Port Power

500 kHz to ≤ 3 GHz

+3 dB

–25 dBm

3 GHz to ≤ 6 GHz

0 dB

–25 dBm

Frequency Range

Dynamic Range

2 MHz to ≤ 4 GHz

100 dB

4 GHz to ≤ 6 GHz

90 dB

Transmission Dynamic Range

The transmission dynamic range (the difference between

test port power and noise floor) using 10 Hz IF Bandwidth

and High Port Power is shown in the following table.

Block Diagram

As shown in the following block diagram, the LMR Master

has a 2-port, 1-path architecture that automatically

measures 2 S-parameters with error-correction precision

inherent to VNA operation.

DUT

Receiver

Port 1

Reference

Receiver

Port 2

Port 1

Bridge/Coupler

Source

LO

S

11

S

21

The above illustration is a simplified block diagram of LMR Master’s

2-port, 1-path architecture. The magnitude and phase information

gained from vector network data enables the LMR Master to

make significant error corrections and provide improved field

measurements.

Typical Test Port Power

LMR Master supports selection of either High (default)

or Low test port power. Changing power after calibration

can degrade the calibrated performance. Typical power by

bands is shown in the following table.

Typical Sweep Speed

The typical sweep speed for IF Bandwidth of 100 Hz,

1001 data points, and single display is shown in the

following table. The two receiver architecture will

simultaneously collect S

21

and S

11

(or S

12

and S

22

)

in a single sweep.

Frequency Range

Typical Sweep Speed

500 kHz to 6 GHz

850 μs / point

LMR Master™ S412E Specifications

Vector Network Analyzer

Page 4 of 24

Advertising