Unless otherwise stated, WAN interfaces provide two input ports and two

output ports allowing any one of the following connection configurations to

be used:

•Terminal Mode - uses one input and one output

•Bridged Monitor Mode - uses both inputs and outputs for bi-directional

monitoring at test access points; a high input impedance should normally be

selected

•Jack Monitor Mode – uses both inputs only for bi-directional monitoring at

protected monitor points

•Through Monitor Mode (Drop and Insert Mode) - uses both inputs and

outputs providing bi-directional repeater and monitoring functionality. Each

receiver input is terminated. The insert mode can place test traffic like SLA,

PING, LMI emulation, Traffic Generation and Simulation.

The operation of the physical interface is often critical in determining the

cause of network problems. Therefore, the Advisor tracks errors at the

physical layer, and it stamps the information with a 1 ms time stamp. Signal

events are recorded on the display for both the line (network) side as well as

the equipment (user/subscriber) side. The time of the last occurrence of a

particular event is recorded as well. Events are saved in the buffer and can be

logged to disk.

Line status is displayed in real time. All of the events listed in the “Physical

layer alarms and statistics” section for each interface below are saved in the

buffer and counted in the line status display. These events may be logged to

disk. Critical parameters marked with an asterisk (*) are also displayed in

large green or red boxes in the line status display for easy “at-a-glance”

viewing.

T

T

T

T

T1

11

11/////DS

DS

DS

DS

DS1

11

11 (((((1

11

11.....5

55

554

44

44 4

44

44 M

M

M

M

Mb

b

b

b

b/////s)

s)

s)

s)

s) J

J

J

J

J2

22

22 2

22

229

99

99 8

88

88B

B

B

B

B a

a

a

a

an

n

n

n

nd

d

d

d

d J

J

J

J

J2

22

22 2

22

229

99

99 9

99

99B

B

B

B

B

General

This interface comprises a choice of two plug-in modules (J2298B and J2299B)

suitable for plugging into the J2300C/D Advisor mainframe and the J2900A

High Speed undercradle. The J2298B has RJ-48C connectors and the J2299B

has RJ-45 connectors. These modules handle cell and frame-based

technologies, i.e. ATM, Frame Relay, ISDN, HDLC, X.25, PPP, and BERT. These

interfaces do not support Frame Relay CIR and SLA measurements.

W

W

W

W

WA

A

A

A

AN I

N I

N I

N I

N Int

nt

nt

nt

nte

e

e

e

errrrrfffffa

a

a

a

ac

c

c

c

ce Sp

e Sp

e Sp

e Sp

e Spec

ec

ec

ec

ecif

if

if

if

ific

ic

ic

ic

ica

a

a

a

atttttions

ions

ions

ions

ions

Common to Inputs and Outputs:

Connectors:

J2298B: balanced 100 Ohm RJ-48C and WECO mini-Bantams

J2299B: balanced 100 Ohm RJ-45 and WECO mini-Bantams

Line Code:

ATM: B8ZS

WAN: B8ZS, AMI

Framing:

ATM:

Extended Super Frame (ESF)

WAN:

Extended Super Frame (ESF)

D4 (Super Frame) Ft and Fs, or Fs only

Fractional, any multiple of 56 kbps or 64 kbps channels

Unframed 1.544 Mbps

Cell Mapping:

Direct (normal mode, ITU-T G.804) and PLCP (obsolete mode)

Cell Scrambling:

Conforms to ITU-T I.432.1 (x