Atec Agilent-E7495A User Manual

E7495a/b base station test set specifications

Advertising
background image

Chapter 21

625

Safety and Regulatory Information and Specifications

E7495A/B Base Station Test Set Specifications

Safety and Regula

tory In

fo

rmatio

n

a

n

d

Sp
ecification

s

E7495A/B Base Station Test Set Specifications

Specifications describe the test set's warranted performance and are valid over the
entire operating/environmental range unless otherwise noted.

The E7495A/B Base Station Test Set has been tested to the specifications of IEC
529 “Degrees of Protection Provided by Enclosures (IP Code).” Using this standard,
the E7495A/B is rated IPX4.

Supplemental Characteristics are intended to provide additional information useful
in applying the test set by giving typical, but non-warranted performance
parameters.

Characteristics and specifications are shown as follows:

Bold type indicates a warranted, hard specification.

Normal type indicates a nominal value. Nominal values are design center values
and not normally tested during manufacturing.

Supplemental characteristics are intended to provide additional information useful
in applying the instrument by giving typical, but non-warranted, performance
parameters. These characteristics are shown in italics or labeled as “typical,” or
“usable to”.

Table 21-1

E7495A/B Base Station Test Set General Specifications

Unless otherwise noted, the following specifications apply to all measurements/tools using Port 2.

Frequency accuracy:

Using internal time base:
Internal frequency aging
With GPS lock for

> 15 minutes:

± 1 ppm

± 1 ppm aging/year

± 0.03 ppm

Input frequency range:

500 kHz to 2700 MHz
(Refer to individual measurement tools below for applicable
frequency ranges.)

Maximum input level:

+ 20 dBm (.1 W), + 50 dBm with supplied attenuator

Advertising