Specifications

Mainframe Specifications

HORIZONTAL SYSTEM (Time Base)
Scale Factor
(Full scale is ten divisions.)
Minimum

2 ps/div (with 86107A: 500 fs/div)

Maximum

1 s/div

Delay

1

Minimum

≥24 ns (Standard), 28 ns (Option 001)

Maximum

1000 screen diameters or 10 s, whichever is smaller

Time Interval Accuracy

≤8 ps + 0.1% of reading (dual marker measurement performed at a temperature within

±5°C of horizontal calibration temperature)

8 ps + 0.1% + 0.5 ps/°C to (5 < absolute temperature value <15°C)

Time Interval Resolution

≤ (screen diameter)/(record length) or 62.5 fs, whichever is larger

Display Units

Bits or time (TDR mode–meters)

VERTICAL SYSTEM (Channels)
Number of Channels

4 (simultaneous acquisition)

Vertical Resolution

12 bit A/D converter (up to 15 bits with averaging)

Full Resolution Channel Scales

Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution

from the front panel knob

Adjustments

Scale, offset, activate filter, sampler bandwidth, attenuation factor, transducer conversion factors

Typical Acquisition Rate

40 kHz per Channel

Record Length

Manual setting between 450 and 4050 samples (increments of 1, 1350 default sample setting)

or using up arrow/down arrow keys, increments of 450 samples: 450, 900, 1350 (default) etc.
up to 4050 samples

8

Specifications describe warranted performance over the temperature range of +10°C to + 40°C (unless otherwise noted). The specifications are
applicable for the temperature after the instrument is turned on for one (1) hour, and while self-calibration is valid. Many performance parameters are
enhanced through frequent, simple user calibrations. Characteristics provide useful, non-warranted information about the functions and
performance of the instrument. Characteristics are printed in italic typeface.

Factory Calibration Cycle -For optimum performance, the instrument should have a complete verification of specifications once every twelve (12) months.

General Specifications

This instrument meets Agilent Technologies’ environmental specifications (section 750) for class B-1 products with exception as described for
temperature and condensation. Contact your local field engineer for complete details.

Temperature
Operating

10°C to +40°C (50°F to + 104°F)

Non-operating

–40°C to +70°C (–40°F to +158°F)

Humidity
Operating

Up to 90% humidity (non-condensing) at +40°C (+104°F)

Non-operating

Up to 95% relative humidity at +65°C (+149°F)

Altitude
Operating

Up to 4,600 meters (15,000 ft)

Non-operating

Up to 15,300 meters (50,000 ft)

Vibration
Operating

Random vibration 5–500 Hz, 10 minutes per axis, 2.41 g (rms)

Non-operating

Random vibration 5–500 Hz, 10 minutes per axis, 0.3 g (rms); Resonant search, 5–500 Hz

swept sine, 1 octave/min sweep rate, 0.75 g, 5 minute resonant dwell at 4 resonances/raxis

Power Requirements
Voltage

90–132 or 198–264 Vac, 48–66 Hz

Power (including modules)

604 VA; 391 W

Weight
Mainframe without modules

15.5 kg (34 lb)

Each Module

1.2 kg (2.6 lb)

Mainframe Dimensions (excluding handle)
Without front connectors and rear feet

215.1 mm H x 425.5 mm W x 566 mm D (8.47 in x 16.75 in x 22.2 in)

With front connectors and rear feet

215.1 mm H x 425.5 mm W x 629 mm D (8.47 in x 16.75 in x 24.8 in)

1 Time offset relative to the front panel trigger output on the instrument mainframe.