Magnitude characteristics – Atec Agilent-4395A User Manual
Page 4
Magnitude Characteristics
Absolute amplitude accuracy (R, A, B)
at –10 dBm input, input attenuator: 10 dB, frequency
≥
100 Hz, IFBW
≤
3 kHz, 23 ± 5°C, . . . . . . <±1.5 dB
Ratio accuracy (A/R, B/R) (typical for A/B)
at –10 dBm input, input attenuator: 10 dB, IFBW
≤
3 kHz, 23 ± 5°C,
. . . . . . . . . . . . . . . . . . . . . . . <±2 dB
Dynamic accuracy (A/R, B/R) (typical for A/B)
Input Level
(relative to full scale input level –10 dB)
Dynamic Accuracy
1
frequency
≥
100 Hz
0 dB
≥
input level > –10 dB
±0.4 dB
–10 dB
≥
input Level
≥
–60 dB
±0.05 dB
–60 dB > input level
≥
–80 dB
±0.3 dB
–80 dB > input level
≥
–100 dB
±3 dB
1. R input level (B input level for A/B) = full scale input level –10 dB, IFBW = 10 Hz, 23 ± 5°C
Figure 1-1. Magnitude Dynamic Accuracy
4