Magnitude characteristics – Atec Agilent-4395A User Manual

Page 4

Advertising
background image

Magnitude Characteristics

Absolute amplitude accuracy (R, A, B)

at –10 dBm input, input attenuator: 10 dB, frequency

100 Hz, IFBW

3 kHz, 23 ± 5°C, . . . . . . <±1.5 dB

Ratio accuracy (A/R, B/R) (typical for A/B)

at –10 dBm input, input attenuator: 10 dB, IFBW

3 kHz, 23 ± 5°C,

. . . . . . . . . . . . . . . . . . . . . . . <±2 dB

Dynamic accuracy (A/R, B/R) (typical for A/B)

Input Level

(relative to full scale input level –10 dB)

Dynamic Accuracy

1

frequency

100 Hz

0 dB

input level > –10 dB

±0.4 dB

–10 dB

input Level

–60 dB

±0.05 dB

–60 dB > input level

–80 dB

±0.3 dB

–80 dB > input level

–100 dB

±3 dB

1. R input level (B input level for A/B) = full scale input level –10 dB, IFBW = 10 Hz, 23 ± 5°C

Figure 1-1. Magnitude Dynamic Accuracy

4

Advertising