Options, Option 001 (power and dc bias enhancement), Test signal – Atec Agilent-E4980A User Manual

Page 19

Advertising
background image

19

The following options are available for the E4980A LCR Meter.

Option 001 (Power and DC bias enhancement)

Increases test signal voltage and adds the variable DC bias voltage function.

Measurement parameters

The following parameters can be used.

• Lp-Rdc
• Ls-Rdc
• Vdc-Idc

where

Rdc Direct-current

resistance

(DCR)

Vdc Direct-current

voltage

Idc Direct-current

electricity

Test signal

Signal level

Table 30. Test signal voltage

Range

0 Vrms to 20 Vrms (test frequency ≤ 1 MHz)

0 Vrms to 15 Vrms (test frequency > 1 MHz)

Resolution

100 µVrms (0 Vrms - 0.2 Vrms)

200 µVrms (0.2 Vrms - 0.5 Vrms)

500 µVrms (0.5 Vrms - 1 Vrms)

1 mVrms (1 Vrms - 2 Vrms)

2 mVrms (2 Vrms - 5 Vrms)

5 mVrms (5 Vrms - 10 Vrms)

10 mVrms (10 Vrms - 20 Vrms)

Setup accuracy

normal

±(10% + 1 mVrms) (test signal voltage ≤ 2 Vrms)

(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)

±(10% + 10 mVrms) (Test frequency ≤ 300 kHz,

test signal voltage > 2 Vrms) (spec.)

±(15% + 20 mVrms) (test frequency > 300 kHz,

test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,

test frequency > 1 MHz : typ.)

Constant

1

±(6% + 1 mVrms) (test signal voltage ≤ 2 Vrms)

(test frequency ≤ 1 MHz : spec. , test frequency > 1 MHz : typ.)

±(6% + 10 mVrms) (test frequency ≤ 300 kHz,

test signal voltage > 2 Vrms) (spec.)

±(12% + 20 mVrms) (test frequency > 300 kHz,

test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,

test frequency > 1 MHz : typ.)

Options

Note

Option xxx is described as E4980A-xxx in the
order information

1. When auto level control function is on.

Advertising