Atec Dewetron-DEWE-3020 User Manual

Dewe-3020, Additional interfaces and sensors

Advertising
background image

107

Software Components Signal Conditioning Front-ends Instruments Dedicated Instruments DEWE2

www.dewetron.com/dewe-3020

All-In-One

DEWE-3020

Most portable all-in-one instrument
8 isolated DAQP or 16 differential MDAQ analog inputs
Simple expansion to 16 isolated DAQP modules
3 PCI slots for A/D and other cards (1394, ARINC, 1553…)
15.4” TFT wide-screen display with touch-screen

DEWE-3020 series

Input specifications

DEWE-3020

DEWE-3021

Slots for DAQ or PAD modules

8

-

Expansion connector for additional 8 channels

1

-

MDAQ input channels

-

Up to 16

Main system

1)

Total PCI slots

1 full length / 2 half length

3 half length

Hard disk

1000 GB

Data throughput

Typ. 80 MB/s

Power supply

95 to 260 V

AC

Display

15.4“ TFT wide-screen display with touch-screen, 1280 x 800 pixel

Processor

Intel

®

Core™ i5/i7

RAM

3 GB

Ethernet

10/100/1000 BaseT

USB interfaces

3

RS-232 interface

1

Storage drive

Internal DVD +/-RW burner

Operating system

Microsoft

®

WINDOWS

®

7

Dimensions (W x D x H)

377 x 168 x 284 mm (14.8 x 6.6 x 11.2 in.)

Weight

Typ. 8.5 kg (18 lb.)

Typ. 8 kg (17 lb.)

Environmental specifications
Operating temperature

0 to +50 °C, down to -20 °C with prewarmed unit

Storage temperature

-20 to +70 °C

Humidity

10 to 80 % non cond., 5 to 95 % rel. humidity

Vibration

2)

MIL-STD 810F 514.5, procedure I

Shock

2)

MIL-STD 810F 516.5, procedure I

1)

Please find current specifications in the latest price list

2)

Tested with Solid State Disk

Additional interfaces and sensors

Measurements are not limited to just classic analog and digital signals. Please find further detailed information to
expand your system in the chapter “Components”.

Needed to complete the system

DEWE-ORION “A/D Boards” offer simultaneous sampled
analog inputs, synchronous digital I/Os, high-performance
counters and high-speed CAN interfaces. DAQP or MDAQ
signal amplifiers and software are needed as well.

Options to expand the system

Add further “Interface Cards” like ARINC-429, 1553,
PCM telemetry, FireWire

®

and analog output or special

“Sensors” like synchronized video, industrial encoders
(RIE-360) or GPS.

DEWESoft

DAQP/MDAQ

A/D card

DEWE-30-8-EXP

VIDEO

FlexRay

Choose from two models

Add your choice of signal conditioning,
A/D board(s) and software to complete these systems