Atec AEMC-DL-1081_DL-1080 User Manual

Data loggers 8-channel logger, Features, Specifications

Advertising
background image

Technical Assistance (800) 343-1391

FEATURES

8 universal analog input channels

8 digital I/Os (individually configured as

inputs or outputs)

2 relay outputs (NO, NC and common)

RS485 interface

(Modbus master or slave)

24V

dc

output to power up to eight

4 to 20mA transmitters

Ethernet interface for LAN and

internet use

USB-device interface for configuring,

monitoring and download

USB-host interface for logged data

retrieval through a USB flash drive

Up to 32 configurable alarms

Up to 128 virtual channels

Detachable display (optional)

Up to 100 channels can be logged at a

configurable rate

Data downloading through the

configuration software or through the

freely distributed download DLL

4 input channel types: analog, digital,

remoted and virtual

data loggers

8-Channel Logger

Models DL-1080 & DL-1081

Versatile, powerful and cost

effective data loggers handling

analog, digital and other

types of variables with high

resolution and speed

Special

product

features

R

S

4

85

C O MPAT

IB

L

E

D

C INPUTS

D

IGITAL I/O

UN

IVERSAL

ET

HERNET

P

R

O

GR

AMMA

BL

E

SpEciFicATionS

MODELs

DL-1080 & DL-1081

eLectricaL
Internal Memory

2MB

Power

100 to 240V

ac

, 50/60Hz. 20VA (Max)

Analog Channels Input

Impedance

Thermocouples / Pt100 / Pt1000 / mV: > 2M

Ω

mA: 15

Ω + 1.5V

V: 1.1M

Ω

Accuracy

Thermocouples J, K, T, E and N: 0.2% of the span ± 1°C

Thermocouples R, S and B: 0.2% of the span ± 3°C

Pt100, Pt1000, 0 to 20mA, 4 to 20mA, 0 to 20mV,

0 to 50mV, 0 to 60mV, -20 to 20mV, 0 to 5V and

0 to 10V: 0.15% of the span

Excitation Current

Pt100s: 360μA;

Pt1000s: 320μA

Maximum Pt100/Pt1000

Compensated Cable

Resistance

40

Ω

Digital Inputs

Logic level “0”: from 0 to 0.8V

dc

Logic level “1”: from 3 to 30V

dc

Maximum Input Voltage

30V

dc

Input Current @ 30V

dc

(typical)

3mA

Digital Outputs

Maximum output voltage: 30V

dc

Maximum output current: 200mA

Maximum relay current: 3A @ 250V

ac

; 3A @ 30V

dc

storage Rate

from 1ms to 24 hours

Maximum Channel Logged

100

supported Modbus

Commands

Read Coil Status (01h)

Read Holding Registers (03h)

Write Single Coil (05h)

Write Single Register (06h)

Write Multiple Registers (0Fh)

environMentaL
Operating Temperature

32° to 122°F (0 to 50°C)

Relative Humidity

80% up to 85°F (30°C)

For temperatures higher than 85°F (30°C),

decrease 3% per °C

Altitude

< 2000m

Protection

IP20

Inputs rated CAT II

NEW!

IP

Rated

20

Advertising