Module Specifications

HP 34901A, 34902A, 34908A, 34903A, 34904A

Multiplexer

Actuator

Matrix

General

34901A

34902A

34908A

34903A

34904A

Number of Channels

20+2

16

40

20

4x8

2/4 wire

2/4 wire

1 wire

SPDT

2 wire

Connects to Internal DMM

Yes

Yes

Yes

No

No

Scanning Speed

60 ch/s

250 ch/s

60 ch/s

Open/Close Speed

120/s

120/s

70/s

120/s

120/s

Maximum Input

Voltage (dc, ac rms)

300 V

300 V

300 V

300 V

300 V

Current (dc, ac rms)

1 A

50 mA

1 A

1 A

1 A

Power (W, VA)

50 W

2 W

50 W

50 W

50 W

Isolation (ch-ch, ch-earth) dc, ac rms

300 V

300 V

300 V

300 V

300 V

DC Characteristics

Offset

Voltage

[1]

< 3

µ

V

< 6

µ

V

< 3

µ

V

< 3

µ

V

< 3

µ

V

Initial Closed Channel R

[1]

< 1

< 1

< 1

< 0.2

< 1

Isolation (ch-ch, ch-earth)

> 10 G

> 10 G

> 10 G

> 10 G

> 10 G

AC Characteristics

Bandwidth

10 MHz

10 MHz

10 MHz

10 MHz

10 MHz

Ch-Ch Cross Talk (dB)

[2]

10 MHz

-45

-45

-18

[3]

-45

-33

Capacitance HI to LO

< 50 pF

< 50 pF

< 50 pF

< 10 pF

< 50 pF

Capacitance LO to Earth

< 80 pF

< 80 pF

< 80 pF

< 80 pF

< 80 pF

Volt-Hertz Limit

10

8

10

8

10

8

10

8

10

8

Other

T/C Cold Junction Accuracy

[1]

(typical)

0.8

°

C

0.8

°

C

0.8

°

C

Switch Life No Load (typical)

100M

100M

100M

100M

100M

Switch Life Rated Load (typical)

[4]

100k

100k

100k

100k

100k

Temperature Operating

All Modules – 0

°

C to 55

°

C

Temperature Storage

All Modules – -20

°

C to 70

°

C

Humidity (non-condensing)

All Modules – 40

°

C / 80% R.H.

[1] Errors included in the DMM measurement accuracy specifications
[2] 50

source, 50

load

[3] Isolation within channel 1 to 20 or 21 to 40 banks is -40 dB
[4] Applies to resistive loads only