Atec Associated-Research-7520DT-7550DT-7564S4 User Manual

Atec Equipment

Advertising
background image

SPECIFICATIONS

13

Models 7520DT, 7530DT, 7540DT, 7550DT and 7564SA

Functional Specifications

INPUT

Voltage

115/230 VAC 15 , Single Phase, User selection

Frequency

50/60 Hz 5

Fuse

6.3 Amp 250V Slo-Blo for 7550DT and 7564SA
5 Amp 250V Slo-Blo for 75/20/30/40DT

DIELECTRIC WITHSTAND TEST MODE

Output Rating

5 KV @ 40 mA for 7550DT and 7564SA,
5 KV @ 20mA for 75/20/30/40DT

6 KV @ 10 mA DC for 75/30/40/50DT and 7564SA

Output Adjustment

Range:

0 - 5 KV AC

0 - 6 KV DC

Resolution: 1 volt/step

Accuracy:

(2 of setting 5 volts)

Ramp-HI

12mA peak maximum, ON/OFF selectable

Charge-LO

Range:

0.0 - 350.0 A DC or Auto set

HI-Limit AC Range:

0.00 - 40.00 mA for 7550DT and 7564SA,
0.00 - 20.00 mA for 75/20/30/40DT

Resolution: 0.01 mA/step

Accuracy:

(2 of setting 2 counts)

DC Range:

0 - 9999 A

Resolution: 1 A/step

Accuracy:

(2 of setting 2 counts)

LO-Limit AC Range:

0.000 - 9.999 mA

Resolution: 0.001 mA/step

Accuracy:

(2 of setting 2 counts)

DC Range:

0.0 - 999.9 A

Resolution: 0.1 A/step

Accuracy:

(2 of setting 2 counts)

Arc Detection

Range:

1 - 9

Failure Detector

Audible and Visual

Recycled Equipment (410)685-1997 www.recycledequipment.com email: [email protected]