Atec Baker-DX Series User Manual

Page 4

Advertising
background image

Baker Instrument Company, an SKF Group Company

4812 McMurry Avenue, Fort Collins, CO 80525 USA

T: +1 970-282-1200 – +1 800-752-8272 F: +1 970-282-1010

www.bakerinst.com

® SKF is a registered trademark of the SKF Group.

Baker is a trademark of the SKF Group.
All other trademarks are the property of their respective owners.

© SKF Group 2012

The contents of this publication are the copyright of the publisher and may not be reproduced (even extracts) unless prior written

permission is granted. Every care has been taken to ensure the accuracy of the information contained in this publication but no liability can

be accepted for any loss or damage whether direct, indirect or consequential arising out of the use of the information contained herein.

PUB CM/P8 12200 EN · January 2012

Printed in USA on environmentally friendly paper.

skf.com bakerinst.com

Table 2

Testing specifications.

Test

4 kV model

6 kV model

6 kV HO model

12 kV model

12 kV HO model

Resistance

Source voltage, maximum:

3,9 V

3,9 V

3,9 V

3,9 V

3,9 V

Source current:

600 mA

600 mA

600 mA

600 mA

600 mA

10 000 to 200 000 :

3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

100 to 10 000 :

3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

0,2 to 100 :

2% accuracy

2% accuracy

2% accuracy

2% accuracy

2% accuracy

0,002 to 0,2 :

4%, ±1 m accuracy

4%, ±1 m accuracy

4%, ±1 m accuracy

4%, ±1 m accuracy

4%, ±1 m accuracy

Capacitance

Source voltage, maximum:

3,9 V

3,9 V

3,9 V

3,9 V

3,9 V

Source current, maximum:

600 mA

600 mA

600 mA

600 mA

600 mA

Source frequency:

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

0,001 to 2,6 F at 4 000 Hz: 3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

2,6 to 26 F at 4 000 Hz:

5% accuracy

5% accuracy

5% accuracy

5% accuracy

5% accuracy

Inductance

Source voltage, maximum:

3,9 V

3,9 V

3,9 V

3,9 V

3,9 V

Source current, maximum:

600 mA

600 mA

600 mA

600 mA

600 mA

Source frequency:

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

160 to 5 000 mH at 60 Hz:

3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

0,5 to 160 mH at 60 Hz:

2% accuracy

2% accuracy

2% accuracy

2% accuracy

2% accuracy

0,05 to 0,5 mH at 60 Hz:

3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

Impedance

Source voltage, maximum:

3,9 V

3,9 V

3,9 V

3,9 V

3,9 V

Source current, maximum:

600 mA

600 mA

600 mA

600 mA

600 mA

Source frequency:

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

50 to 4 000 Hz

0,15 to 10 000 at 60 Hz:

2% accuracy

2% accuracy

2% accuracy

2% accuracy

2% accuracy

0,01 to 0,15 at 60 Hz:

3% accuracy

3% accuracy

3% accuracy

3% accuracy

3% accuracy

Phase accuracy at 60 Hz:

< 2 deg.

< 2 deg.

< 2 deg.

< 2 deg.

< 2 deg.

DC tests

Voltage accuracy:

3%

3%

–

3%

–

Maximum resistance:

> 500 G

> 500 G

–

> 500 G

–

Accuracy:

5%

5%

–

5%

–

Minimum resistance:

3 M

5 M

–

10 M

–

Maximum output current:

5 mA

3 mA

–

3 mA

–

Over-current trip:

1,4 mA

1,4 mA

–

1,4 mA

–

Surge

Capacitor size (nF):

40

40

100

40

100

Surge energy:

0,32 J

0,72 J

1,8 J

2,88 J

7,2 J

Short circuit current:

280 A

340 A

450 A

600 A

800 A

65 H load voltage:

4 kV

6 kV

6 kV

12 kV

12 kV

Surge voltage accuracy:

11%

11%

11%

11%

11%

Note: Surge voltage accuracy meets/based on Z540 Standard four times measurement uncertainty (calibrated within 2,5%).