Agilent 8542E and 8546A
EMI Test Receivers

Data Sheet

Agilent Technologies 8542E: 9 kHz to 2.9 GHz

8542E EMI Receiver
85422E Receiver RF Section
85420E RF Filter Section

Agilent Technologies 8546A: 9 kHz to 6.5 GHz

8546A EMI Receiver
85462A Receiver RF Section
85460A RF Filter Section

These specifications apply to both EMI receivers (Agilent
8542E and 8546A) and both receiver RF sections (Agilent
85422E and 85462A) except where noted.

Frequency Specifications

Tuning Range

Frequency Reference

Band 1

9 kHz to 50 MHz

Aging

<± 1 x 10

–7

/year

Band 2

20 MHz to 2.9 GHz

Settability

<± 1 x 10

–8

Band 3

1 GHz to 6.5 GHz*

Temperature stability

<± 1 x 10

–8

Bypass

9 kHz to 2.9 GHz (to 6.5 GHz*)

85422E/85462A

9 kHz to 2.9 GHz (to 6.5 GHz*)

Frequency Span Accuracy

Bands 1 and 2

Band 3 and Bypass

Span

10 MHz

±2% of span + 10 Hz

±4% of span

Frequency Readout Accuracy

Span > 10 MHz

±3% of span

±6% of span

±(frequency readout x frequency reference error** +

85422E/85462A

1% of span + 20% of IF bandwidth + span accuracy + 100 Hz)

Span

10 MHz

±2% of span + 10 Hz

Span > 10 MHz

±3% of span

Marker Count Accuracy

Frequency spans

10 MHz

±(marker frequency x

Counter Resolution

frequency reference

Frequency spans

10 MHz

Selectable from 10 Hz to 100 kHz

error** + counter

Frequency spans > 10 MHz

Selectable from 100 Hz to 100 kHz

resolution + 100 Hz)

Sweep Time

Frequency spans > 10 MHz

±(marker frequency x

Range

20 ms to 100 s

frequency reference

Sweep trigger

free run, single, line, video, external

error** + counter
resolution + 1 kHz)

* For 8546A EMI receiver only

** Frequency reference error = (aging rate x period of time since last

adjustment + initial achievable accuracy + temperature stability)

Specifications

All specifications apply over 0 °C to +55 °C. The EMI receiver
will meet its specifications after 2 hours of storage at a con-
stant temperature, within the operating temperature range, 30
minutes after the analyzer is turned on, and after CAL ALL has
been run.