These specifications apply to the Agilent Technologies E7401A,
E7402A, E7403A, E7404A and E7405A EMC analyzers.

Specifications

All specifications apply over 0° C to +55° C unless otherwise
noted and are covered by the product warranty. The analyzer will
meet its specifications when: it’s within the one year calibration
cycle, AUTO ALIGN [ALL] is selected, stored a minimum 2 hours
within the operating temperature range, turned on for at least 5
minutes, and Align Now RF has been run once every 24 hour period.
Typical performance describes the level at which 80% of the units will
meet or exceed with a 95% confidence level over 20 to 30° C, but is
not covered in the product warranty. Characteristics describe
expected product performance levels that are not covered in the
product warranty.

Frequency specifications
Frequency range
E7401A

50

9 kHz to 1.5 GHz

E7402A

9 kHz to 3.0 GHz

dc coupled

30 Hz

6

to 3.0 GHz

ac coupled

100 kHz

6

to 3.0 GHz

E7403A

dc coupled

9 kHz to 6.7 GHz

dc coupled (Option UKB) 30 Hz

6

to 6.7 GHz

ac coupled

100 kHz to 6.7 GHz

Band
0

9 kHz to 3.0 GHz

1

2.85 GHz to 6.7 GHz

E7404A

dc coupled

9 kHz to 13.2 GHz

dc coupled (Option UKB) 30 Hz

6

to 13.2 GHz

ac coupled

100 kHz to 13.2 GHz

Band LO harmonic = N
0

1-

9 kHz to 3.0 GHz

(Option UKB)

30 Hz

6

to 3.0 GHz

1

1-

2.85 GHz to 6.7 GHz

2

2-

6.2 GHz to 13.2 GHz

E7405A

9 kHz to 26.5 GHz

Band LO harmonic = N
0

1-

9 kHz to 3.0 GHz

0

(Option UKB)

30 Hz

6

to 3.6 GHz

1

1-

2.85 GHz to 6.7 GHz

2

2-

6.2 GHz to 13.2 GHz

3

4-

12.8 GHz to 19.2 GHz

4

4-

18.7 GHz to 26.5 GHz

Frequency reference

(Option 1D5)

Aging

±2 x 10

-6

/year

±1 x 10

-7

/year

Temperature stability

±5 x 10

-6

±1 x 10

-8

Settability

±5 x 10

-7

±1 x 10

-8

Frequency readout accuracy
(start, stop, center, marker)

±(frequency indication
x frequency reference error

1

+ span accuracy + 15% of RBW
+ 10 Hz) + 1 Hz x N

4

Marker frequency counter

2

Accuracy

3

±(marker frequency x frequency
reference error

1

+ counter resolution)

Counter Resolution

Selectable from 1 Hz to 100 kHz

Frequency span
Range

0 Hz (zero span), 100 Hz x N

4

to

the range of the spectrum analyzer

Resolution

2 Hz x N

4

Accuracy (> 2000 sweep points)

Sweep type linear

±0.5% of span

Sweep type log

±2% of span (characteristic)