DATA SHEET > UCS 500N7 > 20120518

TECHNICAL DETAILS

RINGWAVE

RINGWAVE MODULE, RWG/N7 (OPTIONAL)

As per ANSI/IEEE C62.41 and
EN/IEC 61000-4-12

Test voltage

250V - 6,000V ± 10%

Voltage

Wave shape (open circuit)

Rise time (first
peak)

0.5us ± 30%

Oscillation
frequency

100kHz ± 10%

Decaying

Peak 2 to peak 1 = 40 - 110%
Peak 3 to peak 2 = 40 - 80%
Peak 4 to peak 3 = 40 - 80%

Current

Wave shape (short circuit)

Rise time

<= 1.0us

Oscillation
frequency

100kHz ± 10%

Source impedance

12ohm and 30ohm

Peak current

As per selected source impedance

Polarity

Positive, negative

TRIGGER CIRCUIT

Release of pulses

Automatic, manual, external

Synchronization

0° - 360°, resolution 1°

Repetition rate

max. 1Hz (1s - 999s)

OUTPUTS

Direct

Via HV-safety lab connectors

Coupling mode

L, N, PE; line to line and line(s) to
ground

DUT supply

AC: 300V/16A; 50/60Hz
DC: 300V/10A

CRO trigger

5V trigger signal for oscilloscope

TEST ROUTINES

Quick Start

On-line adjustable parameters,
easy-to-use

TELECOM SURGE

TSURGE MODULE, TSURGE7 (OPTIONAL)

Test voltage (o.c.)

250V - 7,000V ± 10%

Energy storage
capacitor

20uF

Polarity

Positive, negative, alternating

Counter

1 - 30,000 or endless, selectable

As per ITU and ETSI
recommendations

Front time

10us ± 30%

Pulse duration

700us ± 20%

As per FCC part 68, Pulse B

Front time

9us ± 30%

Pulse duration

720us ± 20%

Output current

6A - 175A (short circuit)

Rise time

5us ± 30%

Pulse duration

320us ± 20%

As per IEC 61000-4-5

Rise time

6.5us ± 30%

Pulse duration

700us ± 20%

Output current

6A - 175A (short circuit)

Rise time

4us ± 20%

Pulse duration

300us ± 20%

TRIGGER CIRCUIT

Trigger of events

Automatic, manual, external

Repetition rate

max. 0.33Hz (3s - 999s)

OUTPUTS

As per ITU

For 2-wire T1/T2 with 25ohm each

As per FCC part 68

For 2-wire T1/T2 with 25ohm each

As per IEC
61000-4-5

For 4-wire T1/T2/T3/T4 with 100ohm
each

For other requirements special
output configurations are available

OPTIONS

CNV 504S1

Coupling network for 4 telecom lines
as per IEC 61000-4-5

CNV 508S1

Coupling network for 8 telecom lines
as per IEC 61000-4-5