Description specification typical – Atec Agilent-N5072A Series User Manual
Page 10
10
Description
Specification
Typical
Resolution
0.05 dB
Level Accuracy
1
At 50 MHz, 0 dBm, Absolute
± 0.65 dB
(Level Flatness)
2
Stepped Sweep Mode
30 to 300 kHz
300 kHz to 8.5 GHz
—
± 1.0 dB
± 1.0 dB
—
Swept Sweep Mode
30 kHz to 7 GHz
7 to 8.5 GHz
± 2.5 dB
± 3.0 dB
Level Linearity
3
(–15 dBm to Maximum Power)
Stepped Sweep Mode
30 to 300 kHz
300 kHz to 8.5 GHz
—
± 0.75 dB
± 0.75 dB
—
Swept Sweep Mode
All frequency
± 1.5 dB
Power Sweep Range
4
65 dB
1. Power calibration using an external power meter improves level accuracy of the test port output power.
2. Level accuracy of other frequency is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated.
3. Level linearity given is relative to 0 dBm.
4. Stop power may be limited by maximum power.
Test port output signal purity
Description
Specification
Typical
Harmonics (2nd or 3rd)
(at 5 dBm)
30 kHz to 2 GHz
2 to 8.5 GHz
< –25 dBc
< –20 dBc
Non-harmonic spurious
(at 5 dBm)
< –30 dBc