11

Test Port Output (Source)

Test port output frequency

Description

Speciication

Typical

Frequency range

Option 245

Option 285

Option 2H5

100 kHz to 4.5 GHz
100 kHz to 8.5 GHz
100 kHz to 18 GHz

Resolution

1 Hz (100 kHz to 6.5 GHz)
2 Hz (6.5 to 13 GHz)
11 Hz (13 to 18 GHz)

Source stability

± 7 ppm (5 to 40 ºC)

CW accuracy

± 7 ppm

Test port output power

Description

Speciication

Typical

Nominal power (preset power)

–5 dBm

Range

100 kHz to 300 kHz

300 kHz to 8.5 GHz

8.5 to 18 GHz

–20 to –5 dBm
–20 to 0 dBm
–15 to –5 dBm

Resolution

0.05 dB

Level accuracy

At 50 MHz, –5 dBm, absolute

(level latness)

1

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 4.34 GHz

4.34 to 8.5 GHz

8.5 to 12 GHz

12 to 18 GHz

± 0.9 dB

± 3.7 dB
± 2.0 dB
± 1.0 dB
± 1.6 dB
± 3.6 dB
± 5.8 dB

Level linearity

2

–10 to –5 dBm, 100 kHz to 300 kHz

–10 to 0 dBm, 300 kHz to 8.5 GHz
–10 to –5 dBm, 8.5 to 18 GHz

–20 to –10 dBm, 100 kHz to 8.5 GHz

–15 to –10 dBm, 8.5 to 18 GHz

± 1.6 dB
± 1.6 dB
± 1.8 dB

± 2.7 dB
± 2.9 dB

1. Level accuracy of other frequencies is taken at

5 dBm, relative to 50 MHz reference unless otherwise stated. Level accuracy includes averaged total (non-)

harmonics power. Its transient factor is not included.

2. Level linearity given is relative to

5 dBm unless otherwise stated. Level linearity includes averaged total (non-) harmonics power. The level accuracy needs

to be taken into account for test port output power level. Its transient factor is not included.