Test port output (source) – Atec Agilent-E5062A User Manual
Page 9
9
Table 1-9
Test port output signal purity
Description
Specification
Supplemental information
Harmonics (2nd or 3rd)
10 MHz to 2 GHz
< -25 dBc (at 5 dBm, typical)
Non-harmonic spurious
10 MHz to 3 GHz
< -30 dBc (at 5 dBm, typical)
Test port output (source)
Table 1-8
Test port output power
Description
Specification
Supplemental information
Level accuracy (at 23°C ±5°C)
1
300 kHz to 3 GHz
±0.8 dB (at 0 dBm, 50 MHz absolute)
±1.0 dB (at 0 dBm, relative
to 50 MHz reference)
Level linearity (at 23°C ±5°C)
300 kHz to 3 GHz
±0.75 dB (at -5 dBm to 10 dBm)
Range (standard)
300 kHz to 3 GHz
-5 dBm to 10 dBm
Range (extended power)
300 kHz to 3 GHz
-45 dBm to 10 dBm (non-harmonics spurious may limit power range)
Sweep range (without extended power range)
300 kHz to 3 GHz
-5 dBm to 10 dBm
Level resolution
0.05 dB
Test port output (source)
1. Level accuracy for 75
Ω analyzers is not a specification for frequencies >2 GHz; it is a characteristic.