Test port output (source) – Atec Agilent-E5062A User Manual

Page 9

Advertising
background image

9

Table 1-9

Test port output signal purity

Description

Specification

Supplemental information

Harmonics (2nd or 3rd)

10 MHz to 2 GHz

< -25 dBc (at 5 dBm, typical)

Non-harmonic spurious

10 MHz to 3 GHz

< -30 dBc (at 5 dBm, typical)

Test port output (source)

Table 1-8

Test port output power

Description

Specification

Supplemental information

Level accuracy (at 23°C ±5°C)

1

300 kHz to 3 GHz

±0.8 dB (at 0 dBm, 50 MHz absolute)

±1.0 dB (at 0 dBm, relative

to 50 MHz reference)

Level linearity (at 23°C ±5°C)

300 kHz to 3 GHz

±0.75 dB (at -5 dBm to 10 dBm)

Range (standard)

300 kHz to 3 GHz

-5 dBm to 10 dBm

Range (extended power)

300 kHz to 3 GHz

-45 dBm to 10 dBm (non-harmonics spurious may limit power range)

Sweep range (without extended power range)

300 kHz to 3 GHz

-5 dBm to 10 dBm

Level resolution

0.05 dB

Test port output (source)

1. Level accuracy for 75

Ω analyzers is not a specification for frequencies >2 GHz; it is a characteristic.

Advertising