Hewlett Packard 3577A 5Hz - 200 MHz

Network Analyzer Specifications

SOURCE

Frequency

Range

5 Hz - 200 MHz

Resolution

0.001 Hz

Amplitude

± 5 X 10

-8

/day, 0 to 55°C

Range

+15 to -49 dBm (1.26 Vrms to 793m Vrms: 2dBV to -62 dBV)
into a 50Ω load

Resolution

0.1 dB

Accuracy

± 1 dB at +15 dBm and 100 kHz. Below +15 dBm, add the greater
of ± 0.02 dB/dB or 0.2 dB

Flatness

1.5 dBp-p from 5 Hz to 200 MHz

Impedance

50Ω; > 20 dB return loss at all levels

RF Output Connector

50Ω Type N female

Sweep Types

Linear, alternate, cw and log frequency; log amplitude

Sweep Time

100 ms/span to 200 ms/span for frequency sweep; 1 ms/step to 16
s/step for amplitude sweep

Sweep Modes

Continuous, single, manual

Trigger Modes

Free run, immediate, line, external

RECEIVER

Full Scale Input Level

-13 dBV from 10 kHz to 200 MHz with internal 20 dB attenuators
ON (0dBm at 50Ω)

INPUT CHARACTERISTICS

Frequency Range

5 Hz - 200 MHz

Inputs

3 (A, B, R)

Input Impedance

Selectable 50Ω with > 25 dB return loss, or 1 MΩ in parallel with
approximately 30 pF

Input Connectors

50Ω Type N female

Resolution Bandwidth

Selectable 1 kHz, 100 Hz, 10 Hz, or 1 Hz

30 kHz - 200 MHz (50Ω)
30 kHz - 20 MHz (1 MΩ)

Sensitivity (Due to noise and
internal crosstalk between
source and receiver inputs)

Resolution
Bandwidth

Internal 20 dB

Attenuator ON

Internal 20 dB

Attenuator OFF

Advertising