20 GSa/s Sample Rate on All Channels at Once!

• The full real-time bandwidth of

up to 6 GHz is supported on
every channel by the 20 GSa/s
sample rate.

• This industry-leading sample

rate produces more accurate
and repeatable measurements,
avoiding measurement error
and signal aliasing due to under
sampling, as shown above.

• The combination of 6 GHz

bandwidth and 20 GSa/s
sample rate on all channels
makes the 54850 series ideal
for designs that include:
PCI-Express, Serial ATA,
Rapid IO, HyperTransport,
InfiniBand, or Gigabit Ethernet.

20 GSa/s provides accurate measurement.

10 GSa/s is not enough.

5 GSa/s is not enough.

Sample rate

Measured rise time

20 GSa/s

89 psec

10 GSa/s

137 psec

5 GSa/s

238 psec

Example for 90 ps rise time input