Infiniium 90000A Series Oscilloscopes
Performance characteristics

Acquisition

Maximum real-time sample rate

91304A/91204A/90804A: 40 GSa/s (4 channels simultaneously)
90604A/90404A/90254A: 20 GSa/s (4 channels simultaneously)

Memory depth per channel
Standard
Option 50M
Option 100
Option 200
Option 500
Option 01G

20 Mpts on 4 channels
50 Mpts on 4 channels (standard on DSA models)
100 Mpts on 4 channels
200 Mpts on 4 channels
500 Mpts on 4 channels
1 Gpts on 4 channels

Maximum acquired time at highest
real-time resolution
Resolution
Standard
Option 50M
Option 100
Option 200
Option 500
Option 01G

91304A/91204A/90804A 90604A/90404A/90254A
25 ps (40 GSa/s)

50 ps (20 GSa/s)

0.5 ms 1.0 ms
1.25 ms

2.5 ms

2.5 ms

5.0 ms

5.0 ms

10.0 ms

12.5 ms

25.0 ms

25.0 ms

50.0 ms

Data transfer speed
Gigabit Ethernet

USB 2.0 hi-speed (device)

Samples:

1 k

64 k

1 M

16 M

32 M

128 M

MSa/s

(Word): 0.1 1.88 9.25 12.00 12.80 12.80

MSa/s

(Byte): 0.11 1.88 12.60 19.70 20.30 22.00

Samples:

1 k

64 k

1 M

16 M

32 M

128 M

MSa/s

(Word): 0.11 1.88 8.34 8.55 9.07 11.38

MSa/s

(Byte): 0.11 1.88 11.60 14.40 14.90 18.10

Sampling modes
Real-time
Real-time with averaging
Real-time with peak detect

Real-time with hi resolution
Equivalent-time

Segmented memory

Successive single-shot acquisitions
Selectable from 2 to 65534
91304A/91204A/90804A: 40 GSa/s
90604A/90404A/90254A: 20 GSa/s
Real-time boxcar averaging reduces random noise and increases resolution
Resolution: 100 fs
Full bandwidth on all 4 channels, 262,144 sample points maximum memory
Captures bursting signals at maximum sample rate without consuming memory during periods of
inactivity
Number of segments:
Up to 131,072 segments (depending on installed memory depth and model number)
Minimum intersegment time:
91304A / 91204A / 90804A: 2.7 μs
90604A / 90404A / 90254A: 2.5 μs
(the time between the end of the previous acquisition and the beginning of the next acquisition)
Maximum number of segments:
Memory depth:

20 M

50 M

100 M 200 M 500 M 1 G

4096

8192

16384 32768 65536 131072

Roll mode

Scrolls sequential waveform points across the display in a right-to-left rolling motion.
Works at sample rates up to 10 MSa/s with a maximum record length of 40MS.

Filters
Sin(x)/x Interpolation

On/off selectable FIR digital filter. Digital signal processing adds points between acquired data
points to enhance measurement accuracy and waveform display quality.