17

Infiniium Performance Characteristics

Vertical: scope channels

DSO8064A, MSO8064A, DSO8104A, MSO8104A

Input channels

DSO8064A/DSO8104A:

4 analog

MSO8064A/MSO8104A: 4 analog + 16 digital

Analog bandwidth @ 50

Ω (–3 dB)*

DSO8064A/MSO8064A: 600 MHz
DSO8104A/MSO8104A: 1 GHz

Calculated rise time

1

@ 50

DSO8064A/MSO8064A: 583 ps
DSO8104A/MSO8104A: 350 ps

Input impedance*

1 M

Ω ± 1% (13 pF typical), 50 Ω ± 1.5%

Sensitivity

2

1 mV/div to 5 V/div (1 M

Ω)

1 mV/div to 1 V/div (50

Ω)

Input coupling

1 M

Ω: AC, DC; 50 Ω:DC

Hardware bandwidth limit

20 MHz

Vertical resolution

3

8 bits,

≥12 bits with averaging

Channel-to-channel isolation

DC to 50 MHz: 50 dB

(any two channels with

>50 MHz to 500 MHz: 40 dB

equal V/div settings)

>500 MHz to 1 GHz: 30 dB

DC gain accuracy*

2,4

± 1.25% of full scale at full resolution channel scale

Maximum Input Voltage*

1 M

150 V RMS or DC, CAT I
± 250 V (DC + AC) in AC coupling

50

5 Vrms, CAT I

Offset range

Vertical sensitivity

Available offset

1 M

1 mV to <10 mV/div

± 2 V

10 mV to <20 mV/div

± 5 V

20 mV to <100 mV/div

± 10 V

100 mV to <1 V/div

± 20 V

1 V to 5 V/div

± 100 V

50

1 mV to <5 mV/div

± 2 V

5 mV to <200 mV/div

± 5 V

200 mV to 1 V/div

± 20 V

Offset accuracy*

2

± (1.25% of channel offset +2% of full scale +1 mV)

Dynamic range

± 8 div from center screen (1 M

Ω)

± 12 div from center screen (50

Ω)

DC voltage measurement accuracy*

2,4

Dual cursor

± [(DC gain accuracy)+(resolution)]

Single cursor

± [(DC gain accuracy)+(offset accuracy)+(resolution/2)]
Example for single cursor accuracy for 70 mV signal, 10 mV/div, 0 offset: Accuracy =
± [1.25% (80 mV) + (1.25% (0) + 2% (80 mV) + 1 mV ) + (0.4%/2) (80 mV)] = ±3.8 mV