Table 60504-1. Specification and Supplemental Characteristics

SPECIFICATIONS

DC Input Rating:

Current:

0 to 120 A

Voltage

: 3 V to 60 V (minimum dc operation from 0 to 2 V for 0 to 10 A)

Power:

600 W at 40

°

C (derated to 450 W at 55

°

C)

A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL

Constant Current Mode:

Ranges:

0 to 12 A; and 0 to 120 A

Accuracy:

(after 30 second wait):

±

0.12% ± 130 mA (both ranges)

Resolution:

3.2 mA (12 A range); 32 mA (120 A range)

Regulation:

10 mA (both ranges)

Temperature Coefficient:

120 ppm/

°

C ± 8 mA/

°

C (both ranges)

Constant Resistance Mode:

Ranges:

0.017 to 0.5

; 0.5

to 500

; and 5

to 5 k

Accuracy:

± 0.8% ± 5 m

with

12 A at input (0.5

range);

± 0.3% ± 18 mS with

6 V at input (500

and 5 k

ranges)

Resolution:

0.14 m

(0.5

range); 0.54 mS (500 k

range); 0.054 mS (50 k

range)

Regulation:

20 mV with remote sensing (0.5

range); 10 mA (500

and 5 k

ranges)

Temperature Coefficient:

800 ppm/

°

C ± 0.2 m

/

°

C (0.5

range);

300ppm/

°

C ± 1.2 mS/

°

C (500

and 5 k

ranges)

Constant Voltage Mode:

Range:

0 to 60 V

Accuracy:

± 0.1% ± 50 mV

Resolution:

16 mV

Regulation:

20 mV (remote sense); 100 mV (local sense)

Temperature Coefficient:

100 ppm/

°

C ± 5 mV/

°

C

Advertising