Atec Agilent-432A User Manual

Agilent 432a analog power meter

Advertising
background image


Agilent 432A Analog Power Meter

Data Sheet





The Agilent 432A provides high accuracy
over a wide temperature range.

By measuring the output voltage of the
thermistor bridges and computing the
corresponding power, even higher
accuracy of ±0.2 percent ±0.5 microW
can be obtained.

Accuracy can be maintained on even the
most sensitive range because the error
due to thermoelectric effect is reduced
to a negligible level.

Power Range: 7 ranges with full-
scale readings of 10, 30 100 and 300
µW 1, 3 and 10 mW; also calibrated
in dBm from -20 dBm to +10 dBm full
scale in 5 dB steps

Noise: Less than 0.25% of full scale
peak (typical)

Response Time: At recorder output,
35 ms time constant (typical)

Fine Zero: Automatic, operated by
front-panel switch

Zero Carryover: Less than 0.50% of
full scale when zeroed on most
sensitive range

Meter: Taut-band suspension,
individually calibrated, mirror-backed
scales. Milliwatt scale more than 100
mm (4.25 in) long

Calibration Factor Control: 13-
position switch normalizes meter
reading to account for thermistor
sensor calibration factor. Range
100% to 88% in 1% steps

Thermistor Sensor: Thermistor
sensors are required for operation of
the 432A.

Power Consumption: 115 to 230
Vac ± 10%, 50 to 400 Hz, 1.5 watts

No thermoelectric error for high
accuracy over a wide dynamic range

Optional long cables for measurment
convenience

1 year standard warranty





Key Specifications

Power Range: 7 ranges with full-scale readings of 10, 30 100 and 300 uW 1, 3 and 10 mW; also

calibrated in dBm from -20 dBm to +10 dBm full scale in 5 dB steps

Noise: Less than 0.25% of full scale peak (typical)

Response Time: At recorder output, 35 ms time constant (typical)

Fine Zero: Automatic, operated by front-panel switch

Zero Carryover: Less than 0.50% of full scale when zeroed on most sensitive range

Meter: Taut-band suspension, individually calibrated, mirror-backed scales. Milliwatt scale more

than 100 mm (4.25 in) long

Calibration Factor Control: 13-position switch normalizes meter reading to account for thermistor

sensor calibration factor. Range 100% to 88% in 1% steps

Thermistor Sensor: Thermistor sensors are required for operation of the 432A

Advertising
Table of contents