3

150 kHz

f

c

<10 MHz

Amplitude Modulation

RATES:

20 Hz to 10 kHz, 150 kHz

f

c

<10 MHz.

20 Hz to 100 kHz, 10 MHz

f

c

1300 MHz.

DEPTH: to 99%.

ACCURACY

2, 3, 6

:

AM Accuracy

Frequency Range

Rates

Depths

±2% of reading

150 kHz – 10 MHz

50 Hz – 10 kHz

5% – 99%

±1 digit

±3% of reading

150 kHz – 10 MHz

20 Hz – 10 kHz

to 99%

±1 digit

±1% of reading

10 MHz – 1300 MHz

50 Hz – 50 kHz

5% – 99%

±1 digit

±3% of reading

10 MHz – 1300 MHz

20 Hz – 100 kHz

to 99%

±1 digit

±1.5% of reading 1300 MHz – 26.5 GHz

50 Hz – 50 kHz

5% – 99%

±1 digit

±3% of reading

10 MHz – 26.5 GHz

20 Hz – 100 kHz

to 99%

±1 digit

For rms detector add ±3% of reading.

FLATNESS

5, 7

:

Flatness

Frequency Range

Rates

Depths

±0.3% of reading

10 MHz – 1300 MHz

90 Hz – 10 kHz

20% – 80%

±1 digit

±0.3% of reading

10 MHz – 26.5 GHz

90 Hz – 10 kHz

20% – 80%

±1 digit

DEMODULATED OUTPUT DISTORTION:

<0.3% THD for

50% depth.

<0.6% THD for

95% depth.

For f

c

>1300 MHz add 0.4% THD.

FM REJECTION (50 Hz to 3 kHz BW)

3

:

FM Rejection

Frequency Range

Rates

Deviations

<0.2% AM

250 kHz – 10 MHz

400 Hz or 1 kHz

<5 kHz

peak

<0.2% AM

10 MHz – 1300 MHz

400 Hz or 1 kHz

<50 kHz

peak

<0.2% AM

10 MHz – 26.5 GHz

400 Hz or 1 kHz

<50 kHz

peak

RESIDUAL AM (50 Hz to 3 kHz BW): <0.01%

rms

.

Supplemental Characteristics:

DETECTORS: +peak, –peak, ±peak/2, peak hold, average (rms
sinewave calibrated), rms.

MAXIMUM DEPTH, RESOLUTION, AND MAXIMUM DEMODU-
LATED OUTPUT SENSITIVITY ACROSS AN OPEN CIRCUIT
(600

output impedance)

5

:

Maximum

Maximum Demodulated

Depths

Resolution

Output Sensitivity

0.1%

0.01V / percent

AM

peak

40.0%

0.01%

0.1V / percent

AM

peak

<40.0%

0.001%

(rms detector only)

0.1V / percent

AM

rms

<3.0%

Phase Modulation

RATES:

200 Hz to 10 kHz, 150 kHz

f

c

<10 MHz.

200 Hz to 20 kHz, 10 MHz

f

c

1300 MHz.

200 Hz to 20 kHz, 10 MHz

f

c

26.5 GHz.

ACCURACY

3

:

±4% of reading ±1 digit, 150 kHz

f

c

<10 MHz.

±3% of reading ±1 digit, 10 MHz

f

c

1300 MHz.

±3% of reading ±1 digit, 10 MHz

f

c

26.5 GHz.

For rms detector add ±3% of reading.

DEMODULATED OUTPUT DISTORTION: <0.1% THD.

AM REJECTION (for 50% AM at 1 kHz rate)

3

:

<0.03 radians peak (50 Hz to 3 kHz BW).

MAXIMUM DEVIATION, RESOLUTION, AND MAXIMUM
DEMODULATED OUTPUT SENSITIVITY ACROSS AN OPEN
CIRCUIT (600

output impedance)

5

:

Supplemental Characteristics:

MODULATION RATES: usable from 20 Hz to 100 kHz with degraded
performance.

DETECTORS: +peak, – peak, ±peak/2, peak hold, average (rms
sinewave calibrated), rms.

2. Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with rms detector.
3. Peak residuals must be accounted for in peak readings.
5. For optimum flatness, cables should be terminated with their characteristic

impedance.

6. For peak measurements only: AM accuracy may be affected by distortion generat-

ed by the measuring receiver. In the worst case this distortion can decrease accu-
racy by 0.1% of reading for each 0.1% of distortion.

7. Flatness is the variation in indicated AM depth for constant depth on input signal.