background image

10

Power

Supply

Typical operation (watts)

+3.3 VDC

11.2

+5 VDC

22

+12 VDC

5

-12 VDC

5

Total Power

43.2

Environmental

Samples of this product have been

type tested in accordance with the

Agilent Environmental Test Manual

and verified to be robust against the

environmental stresses of Storage,

Transportation, and End-use; those

stresses include but are not limited to

temperature, humidity, shock vibra-

tion, altitude, and power line condi-

tions. Test methods are aligned with

IEC 60068-2 and levels are similar to

MIL-PRF-28800F Class 3.

Operating temperature:

0 to +55 degrees C

Storage temperature:

-20 to +70 degrees C

Relative Humidity:

Type tested:

10 to 90% at 40 degrees C

(non-condensing)

Altitude:

0 to 2000m (6500 ft) above sea level

Safety

Designed for compliance to IEC

61010-1:2001

EMC

Meets the conducted and radiated

emissions and immunity requirements

of IEC 61326:2002 when tested with

EMC shielded filler panels (Agilent

P/N N6030-80007, kit of 6) separating

the controller and the N6030A

module, and in all open slots. The RFI

gaskets must be oriented to the right.

Weight

1.14 kg (2.5 lb)

Security

All user data stored in volatile memory

Dimensions

3U, 4 slot CompactPCI module

8.1 x 13 x 21.6 cm
(3.2 x 5.1 x 8.5 inches)

ISO compliance

This modular instrument is

manufactured in an ISO-900

registered facility in concurrence with

Agilent Technologies, Inc.

commitment to quality.