7

Level accuracy with digital modulation turned on [relative to CW]

Conditions:

[with PRBS modulated data;
if using I/Q inputs,

I

2

+ Q

2

= 0.5 V

rms

, nominal]

1

Level accuracy with ALC on

π/4 DQPSK or QPSK formats
Conditions: With raised cosine or root-raised cosine filter and a

≥ 0.35;

with 10 kHz

≤ symbol rate ≤ 1 MHz; at RF freq ≥ 25 MHz;

power

≤ max specified –3 dB

Option 501-504

Option 506

±0.15 dB

±0.25 dB

Constant amplitude formats [FSK, GMSK, etc]

Option 501-504

Option 506

±0.1 dB

±0.15 dB

Level accuracy with ALC off

1, 2

(±0.15 dB) [relative to ALC on]

Conditions:

After power search is executed, with burst off.

Level switching speed

1

Option 501-504

With Option UNB

Option 506

Normal operation [ALC on]

(< 15 ms)

(< 21 ms)

(< 21 ms)

When using power search manual

(< 83 ms)

(< 95 ms)

(< 95 ms)

When using power search auto

(< 103 ms)

(< 119 ms)

(< 119 ms)

Specifications for Frequency and Power Characteristics

1. Parentheses denote typical performance.
2. When applying external I/Q signals with ALC off, output level will vary directly with I/Q input level.