Atec Agilent-N5173B User Manual

Page 11

Advertising
background image

11

Nonharmonics (CW mode)

1, 2

Range

> 10 kHz offset

Standard (dBc)

9 kHz to < 5 MHz

–65

5 to < 250 MHz

–75

250 to < 750 MHz

–78

750 MHz to < 1.5 GHz

–72

1.5 to < 3 GHz

–66

3 to < 20 GHz

–60

20 to 40 GHz

–54

Subharmonics (CW mode)

9 kHz to 1.5 GHz

None

> 1.5 to 3.2 GHz

–75

> 3.2 to 5 GHz

–67

> 5 to 10 GHz

–67

> 10 to 20 GHz

–56

> 20 to 40 GHz

–53

1. CW mode at +10 dBm.

2. Power line related non-harmonics : 60 Hz to 300 Hz: < 50 dBc. Measured from 1 MHz to 40 GHz.

Jitter

1

(measured)

Carrier frequency

SONET/SDH data rate

rms jitter BW

μUI rms

Picoseconds

155 MHz

155 MB/s

100 Hz to 1.5 MHz

126

0.8

622 MHz

622 MB/s

1 kHz to 5 MHz

62

0.1

2.488 GHz

2488 MB/s

5 kHz to 20 MHz

251

0.1

9.953 GHz

10 kHz to 80 MHz

939

0.094

39.812 GHz

40 kHz to 320 MHz

3408

0.086

1. Calculated from phase noise performance in CW mode at +10 dBm. For other frequencies, data rates, or bandwidths, please consult your sales

representative.