10

Maximum reverse power, nominal

< 1 GHz

50 W

> 1 to < 2 GHz

25 W

> 2 to < 6 GHz

20 W

Max DC voltage

50 VDC

Trip level

2 W

Amplitude switching speed

1

Standard

Option UNZ

Option UNZ, typical

CW mode
SCPI mode

≤ 5 ms, typical

≤ 750 µs

≤ 650 µs

Power search SCPI mode

< 12 ms, measured

List/step sweep mode

≤ 5 ms, typical

≤ 500 µs

≤ 300 µs

Digital modulation on (N5172B only)
SCPI mode

≤ 5 ms, typical

≤ 1.15 ms

≤ 950 µs

Power search SCPI mode

< 12 ms, measured

List/step sweep mode

≤ 5 ms, typical

≤ 900 µs

≤ 400 µs

Alternate power level control (N5172B only)

Switching time

(via waveform markers)

20 µs within ± 1 dB, measured

Functional power range

–15 dBm to –144 dBm, measured

User flatness correction

Number of points

3201

Number of tables

Dependent on available free memory in instrument; 10,000 maximum

Entry modes

USB/LAN direct power meter control, LAN to GPIB and USB to GPIB, remote bus and manual

USB/GPIB power meter control

Sweep modes

See Frequency Specifications section for more detail

1. Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB. Switching speed specifications apply when status register

updates are off.