3

POWER OUTPUT CAPABILITY:
75 VA into 3

exciter or resistive load

CURRENT LIMITING:
Switchable,
Max. 5 A for Vibration Exiter Type 4809
Max. 1.8 A for Mini-Shaker Type 4810

FREQUENCY RESPONSE:
10 Hz to 20 kHz (

±

0.5 dB)

2 Hz to 50 kHz (

±

3 dB)

HARMONIC DISTORTION:
< 0.2% (20 Hz to 10 kHz)
< 0.5% (20 Hz to 20 kHz)
at full output capacity

INPUT IMPEDANCE:
15 k

OUTPUT IMPEDANCE:
< 0.04

(10 Hz to 5 kHz)

< 0.08

(5 kHz to 20 kHz)

PROTECTION:
Short circuit
Excessive heat sink temperature
Input overload

DC STABILITY:
< 25 mV drift for

±

5% supply line variation

< 25 mV drift for ambient temperature variations
between 10 and 40

°

C (50 and 104

°

F)

HUM AND NOISE:
< 5 mV RMS

MAX. VOLTAGE GAIN AT 1 KHZ:
40 dB (

±

1 dB)

ATTENUATOR:
0 to 40 dB in 10 dB steps

GAIN CONTROL:
0 to –× dB logarithmic

POWER REQUIREMENTS:
110, 115, 127, 220 and 240 V AC (

±

5%, 50 to

60 Hz)
Approximately 140 VA
Complies with safety class 1 of IEC 348

DIMENSIONS:
Height:
133 mm (5.2 in)
Width: 210 mm (8.3 in)
Depth: 240 mm (9.5 in)
(KK 0042 Cabinet, 6/12 of 19 in rack module)

WEIGHT:
6 kg (13.2 lb)

Specifications 2706

Ordering Information

Type 2706 Power Amplifier includes the
following accessories:

Mains Cable

JP 0101

Standard Coaxial Plug

3

×

VF 0010

2 A, 250 V Fuse, time-lag

2

×

VF 0013

1 A, 250 V Fuse, time-lag

Brüel&Kjær reserves the right to change specifications and accessories without notice

COMPLIANCE WITH STANDARDS:

CE-mark indicates compliance with: EMC Directive and Low Voltage Directive.

Safety

IEC 348: Safety Requirements for Electronic Apparatus

EMC Emission

EN 50081 –1: Generic emission standard. Part 1: Residential, commercial and
light industry.
EN 50081 –2: Generic emission standard. Part 2: Industrial environment.
CISPR 22: Limits and methods of radio disturbance characteristics of
information technology equipment. Class B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.

EMC Immunity

EN 50082 –1: Generic immunity standard. Part 1: Residential, commercial and
light industry.
EN 50082 –2 : Generic immunity standard. Part 2: Industrial environment.
Note 1: The above is guaranteed using accessories listed in this Product Data
sheet only.
Note 2: RF disturbances above 3 MHz may result in demodulation and overload
conditions.

Temperature

IEC 68 –2 –1 & IEC 68– 2 –2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: +5 to +40

°

C (41 to 104

°

F)

Storage Temperature: –25 to +70

°

C (–13 to 158

°

F)

Humidity

IEC 68 –2 –3: Damp Heat 90% RH (non-condensing at 40

°

C (104

°

F))

Advertising