Tracking generator option (option 0020) – Atec Anritsu-MS2721B User Manual

Page 6

Advertising
background image

Page 6 of 20

Spectrum Master™ MS2721B Spectrum Analyzer Specifications

Tracking Generator Option (Option 0020)

Setup Parameters

Measure Set-up

Off/On, Output Power, Reset Sweep, S21 Insertion loss, Abs Max, Min, Avg (On/Off)

S21 Insertion Loss Set-up

Normalize (Off/On), Rel Reference, Rel Scale, S21 Max, Min, Avg (Off, On), S21 RL Offset

Frequency Range

450 kHz to 7.1 GHz (usable to 100 kHz)

Frequency Resolution

1 Hz

Frequency Accuracy

± 0.3 ppm (25 °C ±25 °C) + aging

Output Power

–40 to 0 dBm

Step Size

0.1 dB nominal

Level Accuracy

± 1.5 dB max (450 kHz to 7.1 GHz, 15 °C to 35 °C)

Zero Span Behavior

CW Output

Output Connector

Type N female, 50 Ω

Damage Level

+ 23 dBm
± 50 VDC (limited dv/dt)
2 kV ESD

DVB-T/H, SFN, BER (Options 0064, 0078, 0057)

See Technical Data Sheet 11410-00436

ISDB-T, SFN (Options 0030, 0032)

See Technical Data Sheet 11410-00436