Trigger characteristics (nom)

General

Trigger source

Internal, external, bus, manual

Trigger count

1 to 1,000,000

Trigger delay

0 s to 3600 s in 1 µs steps

Samples/trigger

1 to 1,000,000

External trigger input (typ)

Connector

Rear panel BNC(f)

Impedance

1 kΩ

Level

TTL compatible

Slope

Selectable positive or negative

Pulse width

> 40 ns min.

Latency

2

Frequency, period: 1 µs + 3 periods

time interval, totalize: 100 ns

External trigger rate

300/s max

1 k/s max

10 k/s max

Damage level

<-5 V, >+10 V

Timebase characteristics (nom)

Timebase reference

Internal, external, or auto

Timebase adjustment
method

Closed-box electronic adjustment

Timebase adjustment
Resolution

10

-10

(10

-11

for Option 010 U-OCXO timebase)

External timebase input (typ)

Impedance

1 kΩ AC coupled

Level (typ)

100 mVrms to 2.5 Vrms

Lock frequencies

10 MHz, 5 MHz, 1 MHz

Lock range

±1 ppm (±0.1 ppm for Option 010 U-OCXO timebase)

Damage level

7 Vrms

Timebase output (typ)

Impedance

50 Ω ± 5% at 10 MHz

Level

0.5 Vrms into a 50 Ω load
1.0 Vrms into a 1 kΩ load

Signal

10 MHz sine wave

Damage level

7 Vrms

1. Continuous, gap-free measurements limits the Gate Time setting to 10 µs to 1000 s in 10 µs steps.
2. Latency does not include delays due to auto-leveling.