ACCURACY–PRESSURE MEASUREMENT

0 to 110% of Full Scale: ±(0.05% of Reading + Floor Term)

Floor Term: 0.005% of FS or display resolution,

whichever is greater

Vacuum: ±(0.25% of Reading

+ Vacuum Floor Term)

Vacuum Floor Term: 0.004 PSI for 16 and 36 PSI

0.01 PSI for 300 PSI

„

„

ACCURACY–MILLIAMPERES

Accuracy: ±(0.025% of Reading + 0.001 mA)

Range: 0 to 55 mA

Resolution: 0.001 mA

Connection: 4 mm jacks, 19 mm (0.75") spacing

„

„

OPERATING TEMPERATURE

0°C to 50°C (32°F to 122°F)

„

„

DISPLAY

Screen: 2 lines of up to 6 digits,

5 mm (0.2") high

Display Rate: 2 readings/second (Adjustable to

1, 3, or 3.7 readings/second)

„

„

POWER

1 x 9V: 90 hours, typical

„

„

DATA/COMMUNICATION

Digital Interface: micro-USB

Includes all effects of operating temperature, linearity, hysteresis,
repeatability and stability, for one year.

All models indicate vacuum

, but vacuum specification applies

to 16 PSI, 36 PSI, and 300 PSI models only.

Not recommended for continuous use at high vacuum.

Includes all effects of operating temperature, linearity, hysteresis,
repeatability and stability, for one year.

mA can be displayed as a percentage, where 0 to 100%
corresponds to either 4 to 20 mA or 10 to 50 mA.

Jacks are compatible with safety sheathed banana plugs.

Non-condensing.

LCD readable in sunlight.

Changes to reading rate requires software: ConfigM30 is a free
download from crystalengineering.net.

Uses one alkaline 9V battery.

Use of USB interface reduces operating time.

Do not use USB in Hazardous locations.

Free Labview drivers available online.

2 LINE
8 CHARACTER
LCD DISPLAY

SENSOR
DIAPHRAGM
LOCATIONS

DIMENSIONS ARE IN MM (INCHES)

137.2

(5.40)

69.8

(2.75)

1/8" FNPT
PRESSURE PORTS

PRESSURE PORT VOLUMETRIC DATA:

0 4mm

SHIELDED

BANANA JACKS

29.2

(1.15)

A

B

LP distance to diaphragm (

A

to

B

): 22.4 (0.88")

LP volume: 2 ml (0.13 cubic inches)

HP distance to diaphragm (

A

to

B

): 17.8 (0.70")

HP volume: 1 ml (0.06 cubic inches)

MICRO-USB

CONNECTION

DO NOT EXCEED MAXIMUMS

30 Series PSI Data Sheet

4534 . C Page 1 of 3

30 Series Calibrator PSI

Advertising