Personal and area monitors, Specifications for indoor models, Alarm operation control input – Atec Narda-8800 Series User Manual

Page 3: T.o.c

Advertising
background image

Personal and Area Monitors

///////////

435 MORELAND ROAD • HAUPPAUGE, NY 11788

USA TEL: 516 231-1700 • INT’L TEL: (1) 516 231-1390 • FAX (1) 516 231-1711

E-MAIL: [email protected] • www.nardaeast.l-3com.com

SPECIFICATIONS FOR INDOOR MODELS

MODEL No.

8810

8815

8820

8825B

Frequency Range

2 to 30 MHz

10 to 500 MHz

0.5 to 18 GHz

0.5 to 100 GHz

Frequency Sensitivity

±1.5 dB

±

2.0 dB

3.0 dB

=

4.5 dB

=

0.5 - 50 GHz

6.0 dB

=

0.5 - 100 GHz

Alarm Threshold

10 % of Standard

>

0.1 mW/cm

1.0 mW/cm

0.5 mW/cm

Overload

CW

Peak

60 mW/cm

60 W/cm

300 mW/cm

100 W/cm

3 dB Intercept Angle

>180

°

>180

°

>180

°

?

>180

°

?

Temperature

Operating

Non-operating

-10

°

C to +55

°

C

-40

°

C to + 75

°

C

Batteries

Circuit

@

Fail-safe

A

9V lithium (supplied)

3V lithium (supplied)

Weight (approx)

24 oz. (0.7 kg)

Calibration Frequencies

B

2, 3, 10 and 30 MHz

10, 30, 100, 300,

400 and 500 MHz

8.2

B

GHz

8.2

B

GHz

Accessories Supplied

Operation Manual, J1 and J2 Mating Connectors

Notes:

=

Peak to peak. The threshold can be factory set to provide optimum performance over a specific band of interest.

>

The alarm threshold is proportionate to IEEE C95.1-1991 Standard and varies over the 2 to 30 MHz band. The threshold is 0.9% of

the H field Maximum Permissable Exposure (MPE) from 3 to 30 MHz and equal to 10% of the E field MPE from 2 to 30 Mhz.

(Controlled Environments)

?

The 3 dB intercept angle is >180

°

with the E field tangential to the mounting plane and >165

°

with the E field perpendicular to the

mounting plane.

@

The rated life is two years. It is recommended that this battery be replaced annually during calibration.

A

The fail-safe battery has a rated life of 5 to 10 years. It is recommended that it be checked annually during calibration and replaced

every three years.

B

Other calibration frequencies available for narrow band applications. Contact the factory.

ALARM OPERATION

CONTROL INPUT

CONDITION

AUDIBLE &

VISUAL INDICATIONS

J1

OUTPUT



Normal Operation

Light flashes every 40 sec.

Low

Low Battery

Alarm chirps, light flashes

Low

TestButtonDepressed

Steady sound and light

High

Alarm

Steady sound and light

High

Main Battery Failure

No sound or light

High

Sensor Failure

Steady sound and light

High

CONDITION

J2 INPUT

Normal Operation

Open Circuit

Alarm Disabled

Short Circuit

People Present

!

Open Circuit

No People Present

!

Short Circuit

Notes:



The STATUS signal available at J1 is a voltage source that

provides >2.4 Vdc into a 500 k

load when “High.”

The voltage drops to <0.8 Vdc when ”Low.” The load

impedance can be reduced to 24 k

providing that

the Main Battery Failure signal is not required.

Jack type is 3.5mm miniature phone.

J2 Input: Normal condition is an open circuit.

A short circuit (<100 ohms) disables the SMARTS.

Jack type is 2.5 mm miniature phone.

!

When used with the Model 8808 Personnel Sensor.

$&