Cmu-b96 additional rf generator option, Path 2 for wcdma – Atec R-S-CMU200 User Manual

Page 33

Advertising
background image

Version 09.00, October 2008

R&S

®

CMU200 Universal Radio Communication Tester

33

R&S

®

CMU-B96 additional RF generator option

Path 2 for WCDMA

Standard

3GPP

FDD

non-signaling mode

Frequency range

728 MHz to 768 MHz
869 MHz to 894 MHz
925 MHz to 960 MHz
1475.9 MHz to 1500.9 MHz
1805 MHz to 1880 MHz
1930 MHz to 1990 MHz
2110 MHz to 2170 MHz
2180 MHz to 2200 MHz

WCDMA signaling mode

band 1

2110 MHz to 2170 MHz

band 2

1930 MHz to 1990 MHz

band 3

1805 MHz to 1880 MHz

band 4

2110 MHz to 2155 MHz

band 5

869 MHz to 894 MHz

band 6

875 MHz to 885 MHz

band 8

925 MHz to 960 MHz

band 9

1844.9 MHz to 1879.9 MHz

band 10

2110 MHz to 2170 MHz

band 11

1475.9 MHz to 1500.9 MHz

band 12

728 MHz to 746 MHz

band 13

746 MHz to 756 MHz

band 14

758 MHz to 768 MHz

Frequency range

band S

2180 MHz to 2200 MHz

Frequency resolution

2.5

kHz

Frequency uncertainty

same as timebase,
see base unit specifications

Error vector magnitude (EVM)

global EVM for DL RMC in line with 3GPP
TS 34.121 C3.1 to C3.4 with
DPCH/CPICH = 0 dB

<8 %, rms

Output level range

without R&S

®

CMU-U99

–115 dBm to –82 dBm

RF 1

with R&S

®

CMU-U99

–103 dBm to –70 dBm

RF 2

–103 dBm to –70 dBm

Output level resolution

RF 1 and RF 2

0.1 dB

Output level range

overrange mode

without R&S

®

CMU-U99

–110 dBm to –38 dBm

RF 1

with R&S

®

CMU-U99

–90 dBm to –24 dBm

RF 2

–90 dBm to –24 dBm

RF 3 OUT

–70 dBm to –1 dBm

Output level resolution

for overrange mode

1 dB