Technical specifications, Measurement selection – Atec JDSU-ANT-5 User Manual

Page 9

Advertising
background image

9

ANT-5 SDH ACCESS TESTER

Measurement Selection

The ANT-5 offers direct selection of the following measure-
ment tasks:
– Performance Analysis
– Repetitive BERT
– Automatic Protection Switching (APS)
– OH-Capture (SDH only)
– Delay (RTD)
– Tributary Scan (SDH only)

Performance Analysis

ITU-T Recommendation G.821

ES, EFS, SES, DM, and UAS are evaluated. Pass/fail assessment
is based on line length allocation of 1 to 100%. Evaluation for
higher bit rates (up to 140 Mbps) is obtained using a multi-
plex factor as per annex D of G.821. Measurements can be
made using the following events: bit errors (TSEs), FAS-2,
CRC-4, E bit, code errors (2 Mbps), FAS-34, and FAS-140

ITU-T Recommendation G.826

EB, BBE, ES, EFS, SES, and UAS are evaluated. Pass/fail assess-
ment is based on line length allocation of 1 to 100%.The SES
and UAS thresholds can be set by users.
In-service measurement (ISM)
Simultaneous in-service measurement of the near end and far
end of a selected path. Measurements can be made using the
following events: RSOH B1,MSOH B2,HP B3,FAS-140,FAS-34,
FAS-2, CRC, code errors (2 Mbps), and LP-BIP.

Out-of-service measurement (OOS)

Out-of-service measurement using bit errors in the test pat-
tern (for PDH and SDH).

ITU-T Recommendation G.828 Results

ES, EFS, SES, BBE, SEP, and UAS are evaluated. Pass/fail
assessment is based on path allocation of 1 to 100%.The SES
and UAS thresholds can be set by users.
Hierarchy
RSOH B1, MSOH B2, HP B3, LP-BIP,TSE

ITU-T Recommendation G.829

ES, EFS, SES BBE, and UAS are evaluated.The SES threshold can
be set by users.
Hierarchy
RSOH B1, MSOH B2,TSE

ITU-T Recommendation M.2100

ES, EFS, SES, and UAS are evaluated. Pass/fail assessment is
based on line length allocation
of 1 to 100%.The UAS and BISO (bringing into service objec-
tives) thresholds can be set by users.

PDH systems

Measurements can be made using the following events:
TSE, FAS-1.5, FAS-2, FAS-34, FAS-140, CRC, and code errors
(2 Mbps)

ITU-T Recommendation M.2101

ES, EFS, SES, BBE, SEP, and UAS are evaluated. Pass/fail assess-
ment is based on line length allocation of 1 to 100%.The UAS
and BISO (bringing into service objectives) thresholds can be
set by users. ISMs can be performed simultaneously for the
near end and far end of a selected path.
Measurements can be made using the following events:
TSE, LP-BIP, HP-B3, MSOH-B2, and RSOH-B1

Repetitive BER Test

– BER evaluation over a user-definable period of 1-99 seconds
– Automatically repeating feature
– Progress bar displays the current test period
– Large character display of BER result

Auto Protection Switching (APS)

Operates on SDH and PDH interfaces (2 M)
Trigger criteria

MS-AIS, AU-AIS,TU-AIS, or
bit error service disruption*

Pass/fail time limits

10 to 2000 ms

Resolution

1 ms

*Definition of service disruption:
Measurement starts with any of the following events:
TSE, AIS, LOF, or LOS
Measurement stop trigger

Last event

Overhead Byte Capture (SDH only)

Byte capturing with number and time frame recognition for
linear and ring structures.
Trigger source

K1, K2 byte

Trigger criteria

Manual, Compare, Compare Not

Delay (RTD)

Resolution

±1 µs

Except for:
E1 PDH

±100 µs

E1 SDH VC-12

±100 µs

E2 (within PDH E3 or E4)

±10 µs

VC-11/-12 bulk

±10 µs

Measurement range

10 s

VC-12 Tributary Scan (SDH only)

Enables sequential BER testing of C12 channels using config-
ured test pattern. Automatically scans selected VC-12 contain-
ers for defects and anomalies.

Technical
Specifications-

Measurement Selection

Figure 11: Measurement selection

Figure 12: G.826 performance analysis

Advertising