Display of results, Audio monitor (option upl-b5), Extended analysis functions (option upl-b6) – Atec Rohde-Schwarz-UPL User Manual

Page 22: Hearing aids test accessories (option upl-b7), Modification upl-u3

Advertising
background image

22

Audio Analyzer UPL

Display of results

Units
Level (analog)

V, dBu, dBV, W, dBm,
difference (

), deviation (

%) and

ratio (without dimension, %, dBr)
to reference value

Level (digital)

FS, %FS, dBFS, LSBs
deviation (

%) or ratio (dBr)

to reference value

Distortion

% or dB, referred to signal amplitude,
THD and THD+N in all available level
units (absolute or relative to selectable
reference value)

Frequency

Hz, difference (

), deviation (

%) and

ratio (as quotient f/f

ref

, 1/3 octave,

octave or decade) to reference value
(entered or stored, current generator
frequency)

Phase

°

, rad, difference (

) to reference value

(entered or stored)

Reference value (level):
Fixed value (entered or stored).
Current value of a channel or generator signal: permits direct measurement of
gain, linearity, channel difference, crosstalk. In sweep mode, traces (other
trace or loaded from file) can be used as a reference too.

Graphical display of results
Monitor (not UPL66)

8.4“ LCD, colour

Display modes

– display of any sweep trace
– display of trace groups
– bargraph display with

min./max. values

– spectrum, also as waterfall display
– list of results
– bar charts for THD and

intermodulation measurements

Display functions

– autoscale
– X-axis zoom
– full-screen and part-screen mode
– 2 vertical,1 horizontal cursor line
– search function for max. values
– marker for harmonics (spectrum)
– user-labelling for graphs
– change of unit and scale also

possible for loaded traces

Test reports
Functions

– screen copy to printer, plotter or file

(PCX, HPGL, Postscript)

– lists of results
– sweep lists
– tolerance curves
– list of out-of-tolerance values

– equalizer

traces

Printer driver

supplied for approx. 130 printers

Plotter language

HP-GL

Interfaces

2 x RS-232-C, Centronics,
IEC 625 (option UPL-B4)

Storage functions

– instrument settings, optionally with

measured values and curves

– spectra
– sweep results
– sweep lists
– tolerance curves
– equalizer traces

Remote control

via IEC 625-2 (IEEE 488) and RS-232;
commands largely to SCPI
(option UPL-B4)

Audio monitor (option UPL-B5)

Headphones connector

6.3 mm jack

Output voltage (U

P

)

max.

8

V

Output current (I

P

)

max. 50 mA

Source impedance

10

, short-circuit-proof

Recommended headphone impedance 600

Extended analysis functions (option UPL-B6)

Coherence and transfer functions

can be displayed simultaneously

Frequency range

DC to 21.9 kHz

Frequency resolution

from 5.86 Hz

Averaging

2 to 2048

FFT length

256, 512, 1k, 2k, 4k, 8k points

Rub & buzz measurement

simultaneous measurement of frequency
response, rub & buzz and polarity

1)

Frequency range

10 Hz to 110 kHz

Tracking highpass filter

2 to 20 times fundamental

Lower/upper frequency limit

selectable

Measurement time
(200 Hz to 20 kHz, 200 points log.) 2 s

Multi-sine generator function

extended functions

Mode 1

crest factor or phase of each
component selectable

Mode 2

crest factor selectable

Third octave analysis

for analyzer ANLG 22 kHz
and digital 48 kHz

Number of third octaves

30

Frequency range

22 Hz to 22 kHz

Level accuracy

Center frequency

±

0.2 dB

22 Hz to 22 kHz

±

1.0 dB (IEC 1260, class 0)

Stereo sine

in digital generator only

Frequency range

2 Hz

2)

to 21.9 kHz

Frequency

adjustable for each channel

Phase

0 to 360° (same frequency in both
channels)

Level

adjustable for each channel or
channel ratio 2/1

Sweep parameters

frequency and level of channel 1

Other functions

under development

Hearing aids test accessories (option UPL-B7)

Consisting of acoustic test chamber, acoustic 2 cm³ coupler, various battery
adapters, connecting cables, software for measurements to IEC60118 and
ANSI S3.22
Additionally required

options UPL-B5 and UPL-B10

Modification UPL-U3

Change of source impedance of analog generator to 150

(instead of 200

set as standard) at the factory

1)

With UPL-B29 only in base rate mode.

2)

Fixed frequency independent of clock rate.

Advertising