Atec Fluke-87V User Manual

Fluke 83v and 87v digital multimeters, Detailed specifications, For all detailed specifications

Advertising
background image

Fluke 83V and 87V

Digital Multimeters

Detailed Specifications

For all detailed
specifications:

Accuracy is given as ±([% of reading] + [number
of least significant digits]) at 18 °C to 28 °C, with
relative humidity up to 90 %, for a period of one
year after calibration.

For Model 87 in the 4

1

2

-digit mode, multiply the

number of least significant digits (counts) by 10.
AC conversions are ac-coupled and valid from
3 % to 100 % of range. Model 87 is true-rms
responding. AC crest factor can be up to 3 at full
scale, 6 at half scale. For non-sinusoidal wave
forms add -(2 % Rdg + 2 % full scale) typical, for a
crest factor up to 3.

Fluke 87V ac voltage function specifications (true-rms)

Accuracy

Function

Range

Resolution

45 - 65 Hz

30 - 200 Hz

200 - 440 Hz

440 Hz - 1 kHz

1 - 5 kHz

5 - 20 kHz

1

2,4

600.0 mV

0.1 mV

± (0.7 % + 4)

± (2.0 % + 4)

± (2.0 % + 20)

6.000 V

0.001 V

± (1.0 % + 4)

60.00 V

0.01 V

± (0.7 % + 2)

± (2.0 % + 4)

3

unspecified

600.0 V

0.1 V

1000 V

1 V

unspecified

unspecified

Using low pass filter

± (0.7 % + 2)

± (1.0 % + 4)

+ 1 % + 4

unspecified

unspecified

unspecified

-6 % - 4

5

1

Below 10 % of range, add 6 counts.

2

The Fluke 87V is a true-rms responding meter. When the input leads are shorted together in the ac functions, the meter may display a residual reading between 1
and 30 counts. A 30-count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce a much
larger constant error in later measurements.

3

Frequency range: 1 kHz to 2.5 kHz.

4

A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.

5

Specification increases from -1 % at 200 Hz to -6 % at 440 Hz when filter is in use.

Fluke 83V ac voltage function specifications (average responding rms indicating)

Accuracy

Function

Range

Resolution

50 Hz - 60 Hz

30 Hz - 1 kHz

1 kHz - 5 kHz

1

600.0 mV

0.1 mV

± (0.5 % + 4)

± (1.0 % + 4)

± (2.0 % + 4)

6.000 V

0.001 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)

60.00 V

0.01 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)

600.0 V

0.1 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)

2

1000 V

1 V

± (0.5 % + 2)

± (1.0 % + 4)

unspecified

1

Below a reading of 200 counts, add 10 counts

2

Frequency range: 1 kHz to 2.5 kHz