Atec Narda-EF0391 User Manual

Measuring electric fields from 100 khz to 3 ghz

Advertising
background image

NSTS 0209-E0233B-0.0

1 / 2

Subject to change

E-FIELD PROBE

EF0391

Measuring electric fields
from 100 kHz to 3 GHz


using instruments in the NBM-500 family


General public and occupational field

exposure from broadcasting, telecoms and
industrial equipment

Isotropic (non-directional) measurement

64 dB dynamic range without changing

measurement range

Maximum sensitivity starts at 0.2 V/m


The probe contains three orthogonally arranged dipoles with detector
diodes. The three voltages, corresponding to the spatial components,
are available individually at the probe output. The NBM basic unit
calculates the resulting isotropic field strength.

APPLICATIONS

The probe detects electric fields from 100 kHz to 3 GHz, covering the
fields that occur in broadcasting, telecoms, and industry. The high
sensitivity of 0.2 V/m and excellent linearity make it ideal for measuring
human safety limit values in the general public domain.

PROPERTIES

The probe is designed with mechanical and electrical properties ideal for
field use. The probe head is made of foam material to provide effective
protection for the sensors, while having excellent RF characteristics.
The electric destruction limits for pulsed and continuous wave signals
are several times higher than any of the human safety limit values.

CALIBRATION

The probe is calibrated at several frequencies. The correction values are
stored in an EPROM in the probe and are automatically taken into
account by the NBM instrument. Calibrated accuracy is thus obtained
regardless of the combination of probe and instrument.