Level – Atec Rohde-Schwarz-ESL Series User Manual

Page 5

Advertising
background image

Version 02.00, March 2013

Rohde & Schwarz R&S

®

ESL EMI Test Receiver

5

Level

Display range

displayed noise floor to +20 dBm

Maximum rated input level
DC voltage

50 V

CW RF power

30 dBm (= 1 W)

Peak RF power

36 dBm (= 4 W) <3 s

Max. pulse voltage

150 V

Max. pulse energy

pulse width 10 µs

10 mWs

Intermodulation
Third-order intermodulation

intermodulation-free dynamic range, level 2 × –20 dBm, reference level –10 dBm
f

in

< 30 MHz

>54 dBc (TOI +7 dBm, typ. +12 dBm)

f

in

≥ 30 MHz

>60 dBc (TOI +10 dBm, typ +18 dBm)

Second harmonic intercept (SHI)

20 MHz ≤ f

in

≤ 3 GHz

nominal +35 dBm

1 dB compression of input mixer

0 dB RF attenuation, preamplifier = OFF,
f > 200 MHz

nominal +5 dBm

Displayed average noise level (DANL, analyzer mode)

0 dB RF attenuation, termination 50 Ω, RBW = 1 kHz, VBW = 1 Hz,
sample detector, log scaling, tracking generator OFF, normalized to 1 Hz
frequency

preamplifier = OFF

9 kHz to 1 MHz

<–115 dBm (1 Hz)

1 MHz to 10 MHz

<–120 dBm (1 Hz)

10 MHz to 50 MHz

<–130 dBm (1 Hz)

50 MHz to 3 GHz

<–140 dBm (1 Hz)

3 GHz to 5 GHz

<–136 dBm (1 Hz)

5 GHz to 6 GHz

<–130 dBm (1 Hz)

with R&S

®

FSL-B22 option

frequency

preamplifier = ON

9 kHz to 1 MHz

<–130 dBm (1 Hz)

1 MHz to 10 MHz

<–135 dBm (1 Hz)

10 MHz to 50 MHz

<–145 dBm (1 Hz)

50 MHz to 3 GHz

<–152 dBm (1 Hz)

3 GHz to 5 GHz

<–146 dBm (1 Hz)

5 GHz to 6 GHz

<–140 dBm (1 Hz)

frequency

preamplifier = ON, typical values

500 MHz

–162 dBm (1 Hz)

1 GHz

–160 dBm (1 Hz)

3 GHz

–158 dBm (1 Hz)

6 GHz

–147 dBm (1 Hz)

Advertising