Model pro-esd, Model pro-eft, Model pro-surge – Atec Keytek-EMCPro User Manual

Page 2: Model pro-ring, Model pro-telecom, Surge waveform monitoring, Model pro-hpower

Advertising
background image

Model PRO-ESD

ESD per IEC 61000-4-2 and EN 61000-4-2

Trigger Modes

One shot manual, multi-shot tripod

Repetition Rate

Single shot, 1pps or 20pps

Air Discharge Voltage

500V - 8.8kV ±10%

Contact Discharge Voltage

500V - 4.4kV ±10%

Discharge Capacitor

150pF ±10%

Discharge Resistance

330

±10%

Charging Resistance

50M

- 100M

Polarity

Front panel or software controlled

Shot Counter

1 - 999 discharges

Energy Storage

5.8mJ @ 8.8kV

Model PRO-EFT

EFT per IEC 61000-4-4 Edition 2, EN 61000-4-4 and ANSI C62.41

Voltage Waveform

5/50ns ±30%

Peak Voltage

250V - 4.4kV ±5%

Burst Period

300ms ±10%

Burst Duration

15ms ±20%, for pulse frequencies uo to 5kHz, 0.75ms
above 5kHz

Frequency

1-100kHz, in 0.5kHz steps, ±10%

DC Blocking Capacitor

10nF (internal)

Options

Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT &
surge coupler/decoupler
Model CM-CCL: Capacitive coupling clamp
Model CM-CCLC: Coupling clamp cover
Model EFT-ATTN: EFT attenuator for oscilloscope
monitoring

Model PRO-SURGE

Surge for compliant testing per IEC 61000-4-5, EN 61000-4-5, ANSI C62.41 Category B and
UL 1449

Voltage Waveform

1.2/50µs

Peak Voltage

250V - 6.6kV ±5%, 12

mode

250V - 6.0kV ±5%, 2

mode

Peak Current

125A - 3.3kA ±10%

Additional 10

Resistor

Software selectable

Repetition Rate

Up to 4 per minute

Open-circuit Voltage

Front time: 1.2µs ±30%
Duration: 50 µs ±20%

1

Undershoot:

30%

Short-circuit Current

Front time: 8.0µs ±20%

Duration

*

50µs ±20%

Undershoot

30%

Line sync accuracy

±15%, 50 - 277VAC

Options Model

CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT

& surge coupler/decoupler
Model CM-I/OCD: External 8 line coupler/decoupler
for I/O signal lines
Model CM-I/OCD-HS: High speed I/OCD option for
testing data rates to >100kHz

*

Durations are reduced in 12

mode and when coupling multiple lines to PE

Model PRO-RING

**

Ring Wave Surge per ANSI C62.41 Cat. A, B, and UL 864

Voltage Waveform

100kHz damped cosine

Peak Voltage

250 - 6.6kV ±5%

Repetition Rate

<4/minute at 6kV, faster at lower voltages

Open-circuit Voltage

Rise Time: 0.5µs ±30%

Short-circuit Current

Vp/Ip: 12

± 3

or 30

± 8

software selectable

Options Model

CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT

& surge coupler/decoupler

Model PRO-TELECOM

**

Surge Telecom compliant testing per IEC 61000-4-5, EN 61000-4-5, FCC Part 68, ITU K.17,
K.20, K.21 and ETSI

Voltage Waveform

10/700µs (9/720µs FCC Part 68)

Peak Voltage

250V - 6.6kV ±5%

Peak Current

6.25 - 165A +10/-0%, 40

mode

Repetition Rate

Up to 4 per minute

Open-circuit Voltage

Front time: 7.0µs to 11.7µs
Duration: 576µs to 840µs

Short-circuit Current

Front time: 3.5µs to 6.5µs
Duration: 256µs to 384µs

Options

Model CM-TELCD: External coupler for telecom lines

Surge Waveform Monitoring

Lines Monitored

Monitors are automatically switched to match
generator coupling mode

Open-circuit Voltage

1000:1 ±10%

Short-circuit

200:1 ±7%

Current Attenuation

Model PRO-HPOWER

Power Frequency Magnetic Field for compliant testing per IEC 61000-4-8 and EN 61000-4-8

Field Frequency

50Hz/60Hz

Field Amplitude

0.5 - 4A/m, in 0.25A steps, ±10% (with CM-HCOIL)
up to 100A/m with optional external HPOWER-EXT

AC Source

Internal

Resolution

0.25A minimum

Coil Factor

0.65 to 1.00

Coil Resistance

0.05

maximum

Options Model

CM-HMON: Measurement probe for power

frequency magnetic fields
Model CM-HCOIL: 1m x1m magnetic field coil
Model HPOWER - EXT: External generator for power
frequency magnetic field to 30A/m

**

PRO-TELECOM and PRO-RING can not be installed in same unit.

Advertising