Atec Haefely-PSURGE-30-2 User Manual

Page 2

Advertising
background image

Technical Specifications

PSURGE 30.2 Mainframe

Charging voltage range

3...33kV

Test cabinet

39 x 29 x 32 cm (W x D x H)

Capacitors

9uF / 36uF

HV switch

triggered spark gap

Stored energy

5kJ at max. charging
voltage

Impulse polarity

positive, negative,
alternating

HV output

terminals inside the test
cabinet

Impulse circuit

floating, max. voltage
between common and
ground 500V AC

Power mains

208, 230 or 240V
50/60Hz, 1800VA

External synchronisation

0°....360° in 1° steps


Weights and Dimensions (W x D x H, net weight)

PSURGE 30.2

60 x 100 x 190 cm

approx. 320 kg

Impulse Module PS30-CW

Impulse type

Combination wave

Output impedance

2 Ohm

Voltage front time

1.2us

±

30%

Current front time

8us

±

20%

Voltage decay time

50us

±

20%

Current decay time

20us

±

20%

Voltage undershoot

30%

Current undershoot

30%

Voltage amplitude range

3...30kV

Current amplitude range

1.5...15kA

Impulse repetition

up to 6 impulses per
minute at max. voltage

Impulse Module PS30-8x20

Impulse type

current

Output impedance

0.5 Ohm

Voltage front time

not defined

Current front time

8us

±

20%

Voltage decay time

not defined

Current decay time

20us

±

20%

Undershoot

not defined

Undershoot

20%

Charging voltage range

3...15kV

Current amplitude range

6...30kA

Impulse repetition

up to 2 impulses per
minute at max. current

Impulse Module PS30-10x350

Impulse type

current

Output impedance

12 Ohm

Voltage front time

not defined

Current front time

10us

±

20%

Voltage decay time

not defined

Current decay time

350us

±

20%

Charging voltage range

3...15kV

Current amplitude range

240...1200A

Impulse repetition

up to 2 impulses per
minute at max. current

Impulse Module PS30-10x1000

Impulse type

current

Output impedance

37 Ohm

Voltage front time

not defined

Current front time

10us

±

20%

Voltage decay time

not defined

Current decay time

1000us

±

20%

Charging voltage range

3...15kV

Current amplitude range

80...400A

Impulse repetition

up to 2 impulses per
minute at max. current

Advertising