Standard PACKAGE

The EPOCH LTC package includes:
EP-MCA (U8050397): AC adaptor
CASE-10015-0000 (U8764103):

Transport case
910-266: Instrument user’s manual
EPLTC-HS (U8780043): Hand strap
EPLTC-BAT-L (U8760002): Lithium-ion

rechargeable battery
EPLTC-BAT-AA (U8760001): AA battery

cell tray

Instrument I/O cables

EPLTC-C-USB-A-6 (U8840031):

USB-to-Host PC
EPLTC-C-USB-B-6 (U8840033):

USB-to-Client printer
EPLTC-C-RS232-6 (U8840029): RS-232 cable

(1.83 m/6 ft)
EPLTC-C-VGA-6 (U8840035): VGA cable

(1.83 m/6 ft)

Software Options

EPLTC-PRF (U8140067): Manual PRF control
EPLTC-RANGE (U8140065): Extended range
EPLTC-SWP (U8140093): Tunable square

wave pulser
EPLTC-AWS (U8140063): AWS D1.1/D1.5
EPLTC-API 5UE (U8140140): API 5UE
GAGEVIEWPRO-KIT-USB-A-AB (U8140076):

GageView Pro

Optional Accessories

EPXT-EC: External smart battery charger
EP4/CH (U88140055): Chest harness
EPLTC-RPC (U8764008): Rubber protective

case with pipestand
EPLTC-DP (U8780042): Clear display

protectors (10)

EPOCH LTC Specifications*

General

Dimensions (W x H x D)

128.9 mm x 223.3 mm x 55.1 mm (5.07 in. x 8.79 in. x 2.17 in.)

Weight

960 g (2.12 lb) with Li-ion battery

Keypad

English, international symbols, Japanese, or Chinese

Display languages

English, Spanish, French, German, Italian, Japanese, Chinese, Russian, Norwegian,

Swedish

Transducer connection

LEMO 00

Battery

Rechargeable Lithium-ion. Standard AA battery tray

Battery operating time

Lithium-ion rechargeable battery: 8 hours; AA Li-FeS2: 6 hours

AA NiMH cells: 4 hours; AA alkaline cells: 3 hours

Power compatibility

100/120 VAC, 200/240 VAC, 50-60 Hz

Display type

Color transflective LCD, refresh rate of 60 Hz, user-selectable color schemes and

brightness, two split-screen modes and one full-screen mode

Display dimensions (W x H)

640 pixels x 480 pixels, 75 mm x 56 mm (2.95 in. x 2.21 in.), 94 mm (3.7 in.) diagonal

Pulser

Pulser

Fixed width square wave pulser optimized for general flaw detection applications

Pulse width

Fixed width in base instrument. Adjustable from 25 ns to 5,000 ns (0.1 MHz) with

EPLTC-SWP option

Pulse repetition frequency

(PRF)

Automatically software controlled over the range of 10 Hz to 500 Hz.

Manual PRF control option available

Energy settings

100 V to 400 V in 100 V increments

Damping

50 Ω, 100 Ω, 200 Ω, 400 Ω

Receiver

Gain

0 dB to 110 dB — Two user-defined gain step adjustments and presets

Total instrument bandwidth

0.2 MHz to 26.5 MHz at –3 dB

Digital filter settings

Standard: 0.2 MHz to 10.0 MHz Broadband: 2.0 MHz to 21.5 MHz
High Pass: 8.0 MHz to 26.5 MHz

Rectification

Full wave, positive half wave, negative half wave, RF

System linearity

Horizontal: ±0.2% FSW

Resolution

0.25% FSH, amplifier accuracy ±1 dB

Amplitude measurement

0 to 110% full screen height with 0.25% resolution

Reject

0% to 80% full screen height with visual warning

Measurement rate

Equal to PRF in all modes

Calibration

Automated calibration

Velocity, Zero Offset

Straight Beam (First Backwall or Echo-to-Echo)

Angle Beam (Soundpath or Depth)

Test modes

Pulse Echo, Dual, or Through Transmission

Units

Millimeters, inches, or microseconds

Range

3.36 mm to 6,702 mm (0.132 in. to 263 in.) at 5900 m/s (0.2320 in./ms)

Extended range option

3.36 mm to 13,404 mm (0.132 in. to 527 in.) at 5900 m/s (0.2320 in./ms))

Velocity

635 m/s to 15240 m/s (0.0250 in./µs to 0.6000 in./µs)

Zero offset

0 µs to 750 µs

Display delay

59 mm to maximum range (–2.323 in. to maximum range)

Refracted angle

10° to 85° in 0.1° resolution

Measurements

Measurement display locations 5 locations available (manual or auto selection)

Gate (1, 2)

Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight,

Min./Max. Depth, Min./Max. Amplitude

Echo-to-Echo

Standard Gate 2-Gate 1, Optional IF Gate Tracking

Other measurements

Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for

DGS/AVG, AWS D1.1/D1.5 rating (D), Reject Value

DAC/TVG

Standard

DAC points

Up to 50 points, 110 dB dynamic range

Special DAC modes

Custom DAC (up to 6 curves)

Curved surface correction

Standard OD or Bar correction for Angle Beam measurements

Advertising