Setup, Calibration, Setup speed – Atec Panametrics-Olympus-Omniscan-MX2 User Manual

Page 7: S-scan and a-scan display refresh rate

Advertising
background image

7

Setup

Group Wizard for All Essential Parameters

Material selection, with a database for shear and longitudinal

velocities, and configuration of components for flat or curved

surfaces.

A group copy option in the Wizard for fast creation of

symmetrical two-probe inspections.

Wedge selection from a database of Olympus wedges.

Auto probe detection.

Scanner configuration with offsets, skews, and probe positions.

Wizard guidance for phased array, conventional UT, and

TOFD channels.

Detailed interactive and illustrated help menu for every step in

the Wizard.

Weld Overlay and RayTracing: step-by-step Wizard to

configure the weld geometry.

Setup Speed

OmniScan MX

OmniScan MX2

50% faster

S-Scan and A-Scan Display Refresh Rate

OmniScan MX

OmniScan MX2

300% better

Calibration

Code-Compliant Calibration

The Calibration Wizards ensure that every focal law in every
group is the direct equivalent of a single-channel conventional
flaw detector.

Calibration Wizards

Guide the user step-by-step through Velocity, Wedge Delay,

Sensitivity, TCG, DAC, AWS, and DGS calibrations.

Enable experimental or theoretical sensitivity, and TCG curves

based on two, three, or all beams for a real or interpolated

calibration.

Offer a simple, easy-to-use interface that enables all focal laws

to be visualized simultaneously for a particular calibration

task.

Features an interactive help menu with detailed graphics and

definitions, which is available in each step of each Wizard.

Scan plan adjustment using the Focal Law Wizard.

Sensitivity calibration for a defined section in a sectorial scan.

Sensitivity calibration for all beams in a sectorial scan.

UT parameter adjustment using the touch-screen software keypad.