The Model 740 Programmable Thermometer
has a built-in reference junction and will
linearize any of 7 different thermocouple types.
The Model 740/7402 combination includes
nine channels of thermocouple input. An
I EEE-488 interface port is included for easy
system integration or connection to an IEEE-
488 printer for data logging.

Input Expansion

The Model 740/7402 can be expanded up to 81
i nputs by selecting a Model 705 or 706 Scanner
and configuring these with additional 7402 or
7057A thermocouple scanner cards, each xvith
9 inputs. The control for these expansion units

i s through a serial loop connected by standard

coaxial cable, thereby allowing expansion
without using additional IEEE-488 addresses.

Input Programming

Each input can be selected for any of J, K, T, E,
R, S, or B type thermocouples, or for voltage
readback up to 100mV with 1 uV sensitivity.

Any input can also be turned off, allowing the

i nstrument to skip that channel in the scan

sequences. Open thermocouple conditions are
detected and displayed.

Monitor Temperature
Trends

A single input can be read at

programmable intervals to
monitor temperature trends.
Rates up to 20 readings/second

are available for determining

rapid temperature variations,

and the data can be stored in

t he internal memory. Data can

also be transferred to a com-
puter file over the built-in
lEEE-488 bus.

Locate hot spots and examine time-dependent
thermal characteristics. Scan intervals are
programmable, and the maximum rate of 20
readings/second assures a minimum of time
skew between input samples for monitoring
temperature changes.

A SMART PRINT feature can locate the address

and send data to any IEEE-488 printer for data
l ogging of results in a bench configuration.

Non-Volatile Memory

An internal memory stores up to 100 readings.
Data can be recalled from the front panel or
transferred in a block transfer to a controller.
The Model 740 will also give HI, LO, and
AVERAGE of the data memorv.

Simple, Reliable Calibration

Calibration of the Model 740 is easy-there are
no mechanical adjustments. Simply present
the calibration voltages to the input and tell
the instrument to calibrate. Constants are
stored in non-volatile memory, to maintain the
i nstrument's accuracy.