Coupling materials for transducers – Atec Flexim-F610 User Manual

Page 21

Advertising
background image

TSFLUXUS_F601V1-5-1US_Lus, 2014-02-25

21

Technical Specification

FLUXUS® F601

Coupling Materials for Transducers

Technical Data

normal temperature range

(4th character of transducer order

code = N)

extended temperature range

(4th character of transducer order

code = E)

WaveInjector WI-400

< 212 °F

< 338 °F

< 302 °F

< 392 °F

< 536 °F

536 to 752 °F

coupling compound
type N

coupling compound
type E

coupling compound
type E

coupling compound
type E or H

coupling foil type A
and coupling foil
type VT

coupling foil type B
and coupling foil
type VT

type

order code

ambient temperature

material

remark

°F

coupling compound
type N

990739-1

-22 to +266

mineral grease paste

coupling compound
type E

990739-2

-22 to +392

silicone paste

coupling compound
type H

990739-3

-22 to +482

fluoropolymer paste

coupling foil type A 990739-7

max. 536

lead

coupling foil type B 990739-8

> 536 to 752

silver

coupling foil type VT 990739-0

14 to +392

fluoroelastomer

for transducers with transducer
frequency G, H, K

990739-6

for shear wave transducers with
transducer frequency M, P

990739-14

for shear wave transducers IP68
and Lambwave transducers with
transducer frequency M, P

990739-5

for transducers with transducer
frequency Q

coupling foil not to be used for transducer mounting fixture with magnets

Advertising