Minolta/Land Cyclops 52 Infrared Thermometer

Inst # 20001784


Description

• The Minolta-Land Cyclops 52 is a portable thermometer designed for temperature in the

range 600 to 3000° C.

• Temperature is displayed digitally.
• Digital output is available over RS-232 communication.
• Target is viewed simultaneously with the temperature display.
• Wavelength of operation chosen to give maximum freedom from errors due to emissivity.
• Emissivity compensation in range 0.1 to 1.0.
• Three operational modes – Continuous, Peak, and Valley. In the last two the readings

corresponds to the highest and lowest temperature respectively during the period of
operation.

• Response time 30 ms (Display 0.6 s).
• Accuracy ±0.5% reading.
• Wavelength of operation eliminates the effects of water vapor and carbon dioxide in the

atmosphere and the effects of sunlight on temperature readings.

• Three calculating modes – Maximum, Mean, or Minimum of a series of readings.

Operation Notes

During operation, the Cyclops 52 displays the temperature in the digital display inside the
viewfinder. When the trigger is pressed the unit attempts a reading, the display will indicate this
by illuminating.

When the temperature is below 500°C the display will illuminate but will not show a number. If
the temperature is between 500-600°C the number will be displayed and the illumination will
blink rapidly alerting the user that this range falls outside the calibration. Between 600-3000°C
the temperature will be indicated and the illumination will be constant. The LCD value is held
for 30s after measurement.

The Cyclops 52 operation is detailed in the Cyclops 52/53 manual.